9 |
M 5 7 9 1
Important:
If the user proceeds with a method that requires the DIO to be manually changed in the
SYNC Inputs tab (Method 2 and Method 3), the user must also ensure the DIO settings in
the Output tab to be set to Active Low for BOTH Outputs. A Factory Reset also sets the
Output DIO to Active Low by default.
Figure 7: SYNC Configuration Smart Probe Output set to Active Low
4.2
Thermocouple Connection
The iS2-THB-DTC can accept up to two thermocouples. Refer to the thermocouple connector
diagram below to properly connect your thermocouple sensor to the iServer 2 unit.
Figure 8: iServer 2 unit with Thermocouple connectors
Important:
Currently there is Cold Junction Calibration menu available for B-type thermocouples.
To compensate for the B-type thermocouple wire error, users may configure the Offset
(b) and Gain (m) available in the sensor channel settings menu. Refer to section 6.7.1
Gain and Offset
for more information.