5.3

Swamping Condition

If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter
is taking place, a condition known as “swamping” may occur. Readings of a
“swamped” meter cannot be assumed to be either accurate or linear. Flow must 
be restored to below 110% of maximum meter range. Once flow rates are lowered
to within calibrated range, the swamping condition will end. Operation of the meter
above 110% of maximum calibrated flow may increase recovery time.

6.

MAINTENANCE

6.1

Introduction

It is important that the Mass Flow Meter/Controller is used with clean, filtered
gases only. Liquids may not be metered. Since the RTD sensor consists, in part,
of a small capillary stainless steel tube, it is prone to occlusion due to impedi-
ments or gas crystallization. Other flow passages are also easily obstructed.
Therefore, great care must be exercised to avoid the introduction of any potential
flow impediment. To protect the instrument a 50 micron (FMA1700A/1800A Series
Max Flow 15 L/min) or 60 micron (FMA1700A/1800A Series Max Flow 100 and
200 L/min) filter is built into the inlet of the flow transducer. The filter screen and
the flow paths may require occasional cleaning as described below. There is no
other recommended maintenance required. It is good practice, however, to keep
the meter away from vibration, hot or corrosive environments and excessive RF
or magnetic interference.

If periodic calibrations are required they should be performed by qualified per-
sonnel and calibrating instruments, as described in section 7. It is recommended
that units are returned to OMEGA

7

for repair service and calibration.

CAUTION: TO PROTECT SERVICING PERSONNEL IT IS MANDATORY THAT ANY
INSTRUMENT BEING SERVICED IS COMPLETELY PURGED AND NEUTRALIZED 
OF TOXIC, BACTERIOLOGICALLY INFECTED, CORROSIVE OR RADIOACTIVE 
CONTENTS.

6.2

Flow Path Cleaning

Before attempting any disassembly of the unit for cleaning, try inspecting the 
flow paths by looking into the inlet and outlet ends of the meter for any debris that
may be clogging the flow through the meter. Remove debris as necessary. If the
flow path is not unclogged, then proceed with steps below.

Do not attempt to disassemble the sensor. If blockage of the sensor tube is not alle-
viated by flushing through with cleaning fluids, please return meter for servicing.

NOTE: DISASSEMBLY MAY COMPROMISE CURRENT CALIBRATION.

6.2.1

Restrictor Flow Element (RFE)

The Restrictor Flow Element (RFE) is a precision flow divider inside the trans-
ducer, which splits the inlet gas flow by a preset amount to the sensor and main 
flow paths. The particular RFE used in a given Mass Flow Meter depends on the
gas and flow range of the instrument.

Содержание FMA1800A

Страница 1: ...FMA1700A FMA1800A Mass Flow Meter TM e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide NORWALK CT...

Страница 2: ...rves the right to alter specifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...eadings 5 3 Swamping Condition 6 MAINTENANCE 6 1 Introduction 6 2 Flow Path Cleaning 6 2 1 Restrictor Flow Element RFE 6 2 2 FMA1700A 1800A Series Max Flow 15 L min 7 CALIBRATION PROCEDURES 7 1 Flow C...

Страница 4: ...Scaling 7 4 3 Change Decimal Point 8 TROUBLESHOOTING 8 1 Common Conditions 8 2 Troubleshooting Guide 8 3 Technical Assistance 9 CALIBRATION CONVERSIONS FROM REFERENCE GASES APPENDIX 1 COMPONENT DIAGR...

Страница 5: ...ase contact an OMEGA7 customer service representative and request a Return Authorization Number AR It is mandatory that any equipment returned for servicing be purged and neutral ized of any dangerous...

Страница 6: ...ording to the manufacturer s instructions to one and one quarter turns Avoid over tighten ing which will seriously damage the Restrictor Flow Elements RFE s Using a Helium Leak Detector or other equiv...

Страница 7: ...PLY VOLTAGE LIMIT MAY RESULT IN DEVICE PERMANENT DAMAGE IMPORTANT NOTES In general D Connector numbering patterns are standardized There are how ever some connectors with nonconforming patterns and th...

Страница 8: ...as the LCD readout mounting Special lengths of remote extension wiring up to 9 5 feet 3 meters are available on request 2 3 2 Panel Mounting Readouts Another option for the FMA1700A 1800A Mass Flow Me...

Страница 9: ...to the total flow through the transducer In order to sense the flow in the sensor tube heat flux is introduced at two sec tions of the sensor tube by means of precision wound heater sensor coils Heat...

Страница 10: ...rate for 25 to 100 of full scale flow rate GAS PRESSURE 1000 psig 69 bars FMA1700A 1800A Series Max Flow 15 50 and 100 500 psig 34 5 bars FMA1700A 1800A Series Max Flow 200 500 and 1000 L min Optimum...

Страница 11: ...Series Max Flow 1000 L min 3 4 FNPT fittings or 3 4 compression fittings LCD DISPLAY 3 digit LCD maximum viewable digits 1999 0 5 inch high characters On FMA1700A 1800A aluminum or stainless steel mod...

Страница 12: ...0 43 200 44 500 45 1000 TABLE I FMA1700A 1800A SERIES MAX FLOW 15 L min LOW FLOW MASS FLOW METER CODE mL min N2 02 0 to 10 04 0 to 20 06 0 to 50 08 0 to 100 10 0 to 200 12 0 to 500 CODE L min N2 14 0...

Страница 13: ...see page or leak occurs in to the meter 5 2 Flow Signal Output Readings Initiate a controlled gas flow after warm up The flow signal output can be viewed either on the LCD display remote panel meter d...

Страница 14: ...intenance required It is good practice however to keep the meter away from vibration hot or corrosive environments and excessive RF or magnetic interference If periodic calibrations are required they...

Страница 15: ...S 50 L min 100 L min 200 L min 500 L min or 1000 L min DISASSEMBLY NECESSITATES RE CALIBRATION 7 CALIBRATION PROCEDURES NOTE REMOVAL OF THE FACTORY INSTALLED CALIBRATION SEALS AND OR ANY ADJUSTMENTS M...

Страница 16: ...least four times better collective accuracy than that of the Mass Flow Meter to be calibrated be used Equipment required for calibration includes a flow calibration standard and a certified high sensi...

Страница 17: ...ulated screwdriver adjust the ZERO potentiometer R34 through the access window for 0 VDC or 4 mA respectively at zero flow 13 CAUTION The minimum voltage on 0 5 Vdc output can be in the range of 7 to...

Страница 18: ...arily disconnect the gas source Using the multimeter and the insulated screwdriver adjust the ZERO potentiometer R34 through the access window for 0 VDC or 4 mA respectively at zero flow CAUTION The m...

Страница 19: ...e signal Adjust the setting for potentiometer R38 by using the insu lated screwdriver through the access window until reading is within specification 7 3 7 75 Flow Adjustment Using the flow regulator...

Страница 20: ...ired flow and pressure conditions for a given gas or a mix ture of gases It was carefully packed to prevent damage during shipment Should you feel that the instrument is not functioning properly pleas...

Страница 21: ...ect FMA1700A 1800A transducer from power supply remove the shorting condition or check polarities fuse resets automatically Flush clean or disassemble to remove impediments or replace Flush clean or o...

Страница 22: ...ide of the LCD Display Power supply is bad or polarity is reversed PC board is defective REMEDY Readjust ZERO potentiometer R34 through the access hole see page 13 for details Measure voltage on pins...

Страница 23: ...ns 2 and 3 of the D connector Make sure the load resistance is more than 1000 Ohm Return FMA1700A 1800A to factory for repair Check external connections to pins 8 and 9 of the D connector Make sure th...

Страница 24: ...11 NOTE Calibration accuracy can be affected Return FMA1700A 1800A to factory for repair Flush clean or disassemble to remove impediments or replace the filter screen see 6 2 on page 9 NOTE Calibrati...

Страница 25: ...rom FMA1700A 1800A Remove the aluminum housing on the side of the connection cable Side the LCD assembly out of the aluminum housing Reinstall jumper to the appropriate location on the 8 pin header bl...

Страница 26: ...S FROM REFERENCE GASES The calibration conversion incorporates the K factor The K factor is derived from gas density and coefficient of specific heat For diatomic gases K Kgas Note in the above relati...

Страница 27: ...23 APPENDIX 1 COMPONENTS DIAGRAM FMA1700A 1800A METERING PC BOARD...

Страница 28: ...uoride BrF3 3855 1161 6 108 Bromotrifluoromethane Freon 13 B1 CBrF3 3697 1113 6 644 1 3 Butadiene C4H6 3224 3514 2 413 Butane C4H10 2631 4007 2 593 1 Butene C4H8 2994 3648 2 503 2 Butene C4H8 CIS 324...

Страница 29: ...55 Ethyl Acetylene C4H6 3225 3513 2 413 Ethyl Chloride C2H5Cl 3891 244 2 879 Ethylene C2H4 60 365 1 251 Ethylene Oxide C2H4O 5191 268 1 965 Fluorine F2 9784 1873 1 695 Fluoroform Freon 23 CHF3 4967 17...

Страница 30: ...mide CH3Br 5835 1106 4 236 Methyl Chloride CH3Cl 6299 1926 2 253 Methyl Fluoride CH3F 68 3221 1 518 Methyl Mercaptan CH3SH 5180 2459 2 146 Methyl Trichlorosilane CH3 SiCl3 2499 164 6 669 Molybdenum He...

Страница 31: ...35 1592 6 516 Sulfuryl Fluoride SO2F2 3883 1543 4 562 Tetrafluoroethane Forane 134A CF3CH2F 5096 127 4 224 Tetrafluorohydrazine N2F4 3237 182 4 64 Trichlorofluoromethane Freon 11 CCl3F 3287 1357 6 129...

Страница 32: ...pression Fitting Except Model FMA1700A 1800A Series Max Flow 1000 L min DIMENSION INCH LCD VERSION NO LCD A B C C D D E E F G H 10 L min Tube O Diameter 5 60 1 00 1 00 3 00 5 02 0 69 2 69 4 50 50 L mi...

Страница 33: ...29 PARTS OF THE FLOW METER LEFT AND RIGHT VIEWS FRONT VIEW...

Страница 34: ...NOTES 30...

Страница 35: ...or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment...

Страница 36: ...ne Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial...

Отзывы: