10
Procedure:
Screens shown in this manual represent the displays shown with a 500
psi Gauge. The DPG4000 will prompt the technician for the
appropriate pressure at each calibration point.
Use the Pressure Standard to output 500.00 psi (100%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown in the illustration that
follows.
Use the Pressure Standard to output 437.50 psi (87.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 375.00 psi (75%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show ———-. When the readings are
complete the screen should look as shown below.
PSI
437.50
CAL MODE
PSI
375.00
CAL MODE
PSI
312.50
CAL MODE
PSI
500.00
CAL MODE