5.
Apply 75 mV to input terminals at J4.
Let meter stabilize for 10 seconds then
press MIN to accept. Meter will
display:
0IN
(40 mV zero reading)
6.
Apply 0 mV to input terminals at J4.
Let meter stabilize for 10 seconds then
press MIN to accept. Meter will
display:
40IN
7.
Apply 40 mV to input terminals at J4.
Let meter stabilize for 10 seconds then
press MIN to accept.
8.
Meter will briefly display
rST
indicating the span calibration is saved
in non-volatile memory.