T/C

Type

Degrees C

Range

Accuracy

Degrees F

Range

Accuracy

T/C

Material

ISA/ANSI

Color

J

-200.0 to -180.0

±0.5°

-346.0 to -292.0

±0.9°

+Iron

-Connstantan

Jacket

White

Red

Black

-180.0 to -50.0

±0.4°

-292.0 to -58.0

±0.7°

-50.0 to 500.0

±0.3°

-58.0 to 932.0

±0.5°

500.0 to 1200,0

±0.4°

932.0 to 2192.0

±0..7°

K

-230.0 to -100.0

±0.8°

-382.0 to -148.0

±1.4°

+ Chromel®

-Alumel®

Jacket

Yellow

Red

Yellow

-100.0 to 1050.0

±0.4°

-148.0 to 1922.0

±0.7°

1050.0 to 1371.1

±0.5°

1922.0 to 2500.0

±0.9°

T

-260.0 to -200.0

±1.2°

-436.0 to -328.0

±2.2°

+Copper

-Constantan

Jacket

Blue

Red

Blue

-200.0 to -50.0

±0.7°

-328.0 to -58.0

±1.3°

-50.0 to 0.0

±0.4°

-58.0 to 32.0

±0.7°

0.0 to 400.0

±0.3°

32.0 to 752.0

±0.5°

E

-240.0 to -200.0

±0.6°

-400.0 to -328.0

±1.1°

+Chromel

-Constantan

Jacket

Purple

Red

Purple

-200.0 to -100.0

±0.4°

-328.0 to -148.0

±0.7°

-100.0 to 850.0

±0.3°

-148.0 to 1562.0

±0.5°

850.0 to 1000.0

±0.4°

1562.0 to 1832.0

±0.7°

R

-13.3 to 250.0

±1.4°

-1.0 to 482.0

±2.5°

+Pt/13Rh

-Platinum

Jacket

Black

Red

Green

250.0 to 750.0

±0.8°

482.0 to 1382.0

±1.4°

750.0 to 1600.0

±0.7°

1382.0 to 2192.0

±1.3°

1600.0 to 1767.8

±0.8°

2192.0 to 3214.0

±1.4°

S

-18.3 to 100.0

±1.4°

-1.0 to 212.0

±2.5°

+Pt/10Rh

-Platinum

Jackrt

Black

Red

Green

100.0 to 400.0

±1.0°

212.0 to 752.0

±1.8°

400.0 to 1700.0

±0.8°

752.0 to 3092.0

±1.4°

1700.0 to 1767.8

±0.9°

3092.0 to 3214.0

±1.6°

B

315.6 to 550.0

±2.0°

600 to 1022.0

±3.6°

+Pt/30Rh

-Pt/6Rh

Jacket

Grey

Red

Grey

550.0 to 900.0

±1.3°

1022.0 to 1652.0

±2.3°

900.0 to 1150.0

±0.9°

1652.0 to 2102.0

±1.6°

1150.0 to 1820.0

±0.8°

2102.0 to 3308.0

±1.5°

Ranges & Accuracies

 

(Cold Junction Accuracy not included)

Page 3

Configuration

Configure the Calibrator

Move 

POWER  SWITCH  to 

“SOURCE”

  or 

“READ”

Setup

Double click the 

 DIAL KNOB at any time the 

unit is on and the following displays will appear 
for 15 seconds:

Turn  the 

  DIAL  KNOB  to  move  through  the 

menu.  Press  the 

  DIAL  KNOB  to  toggle 

between  OFF  and  ON  or  to  scroll  through  the 
settings. 

MODEL 51#B  V#.##

DOUBLE CLICK

DIAL KNOB

FOR CONFIGURATION

>  EXIT   

 

 

      15

 

TEMP UNITS        °C

 

T/C   

 

 

       K

 

AUTO OFF   

     ON 

EXIT MENU

  -  exits  this  menu  immediately  and 

saves any changes. Menu will automatically exit 
after 15 seconds of inactivity (countdown timer is 
displayed).

TEMP  UNITS

  -  pressing  the  knob  will  toggle 

between °C and °F.

T/C

 - 

520:  pressing  the  knob  will  toggle  between  the 
factory configured T/C Type and mV.

521:  pressing  the  knob  will  cycle  through  T/C 
types J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U 
(T-DIN), Platinel II and mV.

AUTO  OFF

  -  If AUTO  OFF  is  ON,  the  unit  will 

turn  off  after  30  minutes  of  inactivity  to  save 
battery life. If AUTO OFF is OFF the unit will stay 
on until the POWER SWITCH is moved to the off 
position.

Note:

 All settings are remembered even with the 

power  off.  Removing  the  batteries  resets  the 
values to factory defaults.

Simulating  thermocouples  requires  the  use  of 
thermocouple or extension grade thermocouple 
wire. 

Plug  thermocouple  wires  into  the  female 
miniature  thermocouple  connector  mounted  in 
the top end of the housing.

The  OMEGA  CL540A  has  two  banana  jacks 
mounted  in  the  top  end  of  the  housing.  These 
are not temperature compensated and are to be 
used only for millivolt signals.

Connections

Push & Hold

to

Store/Ramp

Push & Turn

for

Fast Dialing

Double Click

for Options

HI

TYPE K °C

Thermocouple Wire

Device

Requiring

Calibration

HI

TYPE K °

C

OFF

THERMOCOUPLE SOURCE

SOURCE

HI

SET

LO

J

T

E

K

R

S

B

N

G

C

D

L

U

P

mV

TYPES

322.1

(Optional Boot Shown)

CL510AZ