Sensor
U.U.T.
DMM
Step
Select
Output
Adjust
Reading*
1
K
32.0°F
P1
0.000 mV ±1µV
2
K
2500.0°F
P41
54.856 mV ±1µV
3
Repeat Steps 1 & 2 As Required
4
J
32.0°F
P1
0.000 mV ±1µV
5
T
32.0°F
P1
0.000 mV ±1µV
6
E
32.0°F
P1
0.000 mV ±1µV
7
N
32.0°F
P1
0.000 mV ±1µV
8
B
1000.0°F
P1
1.439 mV ±1µV
9
R
32.0°F
P1
0.000 mV ±1µV
10
S
32.0°F
P1
0.000 mV ±1µV
11
**
G**
600.0°F
P1
2.238 mV ±1µV
12
C
32.0°F
P1
0.000 mV ±1µV
13
D
32.0°F
P1
0.000 mV ±1µV
Table 2: Calibrator-Mode Zero & Gain Adjustments
*
* exclusive of noise.
** on EC models, G replaced by L. Calibrate L per J.
Sensor
U.U.T.
U.U.T.
Step
Select
Input
Adjust
Reading*
1
K
0.000 mV
P1
32.0 ±0.1°F
2
K
54.856 mV
P11
2500.0 ±0.1°F
3
Repeat Steps 1 & 2 As Required
4
J
0.000 mV
P1
32.0 ±0.1°F
5
T
0.000 mV
P1
32.0 ±0.1°F
6
E
0.000 mV
P1
32.0 ±0.1°F
7
N
0.000 mV
P1
32.0 ±0.1°F
8
B
1.439 mV
P1
1000.0 ±0.5°F
9
R
0.000 mV
P1
32.0 ±0.5°F
10
S
0.000 mV
P1
32.0 ±0.5°F
11
**
G**
2.238 mV
P1
600.0 ±0.2 °F
12
C
0.000 mV
P1
32.0 ±0.2°F
13
D
0.000 mV
P1
32.0 ±0.2°F
Table 1: Meter-Mode Zero & Gain Adjustments
*
* exclusive of noise.
** on EC models, G replaced by L. Calibrate L per J.
26