a)  Adjusting the switching point.

Case 1                             or Case 2

Press and hold the 

SP

button. The display

will show the previous switching pressure
setting, and the dotted bar will flash while
the button is pressed down (case 1).

You can now use the cursor keys to adjust
the switching point upwards or downwards. 
If a cursor key is held down, the values will
change faster. When the cursor key is
released again, the switch-on pressure
setting will cease to change. This setting is
stored and activated when the 

SP

button is

released, after which the display will show
the current pressure value and the bar will 
quit flashing.

b) Adjusting the reset point.

Case 3                             or Case 4

Press and hold the 

RP

button. The display

will show the previous reset pressure
setting, and the dotted bar will flash while
the button is pressed down.

You can now use the cursor keys to adjust
the reset point in the same manner as
described above.

During both adjustment operations, it may
occur that the hysteresis graph changes
from one state to another at the time a
transition is made through the point
"Switching pressure = Reset pressure".
When both points are correctly set, the
hysteresis graph will also be correct. You
can change between 

SP

and 

RP

as often as

you wish until the settings are correct.

c)  Setting a buffering time.

In order to prevent brief pressure “spikes”
or “surges” from causing undesired
switching, a buffering time can be entered.
The effect of this is that pressure changes
are evaluated only if the pressure signal in
question is present for longer than the
buffering time. In order to set a buffering
time, press the button 

SP

before the power

supply is switched on. Release this button
again after the power supply has been
switched on. The display will then show the
buffering time in milliseconds (e.g. 03) or
seconds. The cursor buttons 

can be

used to set the buffering time to 03, 06,
12, 24 or 50 ms or 0.1, 0.2 or 0.4
seconds. When this has been done, press

SP

to store the setting.

d)  Setting the pressure switch to ambient
pressure = 0.

Since ambient pressure varies according to
altitude, the user may re-calibrate the zero
point to match local conditions.

Press the button 

RP

before the power

supply is switched on. Release this button
again after the power supply has been
switched on and the display test has run.
The display will then show "OFS". The
cursor buttons 

can be used to set the

pressure display to 0. When this has been
done, press 

SP

to store the setting.

e)  Hysteresis mode

If it is desired to operate with a fixed
hysteresis value instead of the reset point,
this value can be selected as desired.

In order to set a hysteresis value, the two
buttons 

SP

and 

must be pressed

simultaneously before the power supply 
is switched on.

Release these buttons again after the
power supply has been switched on and the
display test has run. The display will then
show the operating mode. The cursor
buttons 

,

can now be used to change

the operating mode until "HYS" appears in
the display. When this has been done,
press 

SP

to store the setting.

The 

SP

button can be used to display the

switching-point setting, which can be
modified by means of the cursor buttons

,

.

The button 

RP

can be used to display the

hysteresis setting, which can also be
modified by means of the cursor buttons

,

.

Negative hysteresis means: Signal “on”
with rising pressure (case 1).

Positive hysteresis means: Signal “on” with
falling pressure (case 2).

If the switching point is modified, this will
automatically also result in a change in the
reset point by a value equal to the
hysteresis setting.

Case 1

Case 2

f)  Window mode

If it is desired to monitor whether the pressure
lies within a certain range, a switching window
can be created for this purpose. The pressure
switch will then indicate cases in which the
actual pressure lies above or below this area.
In order to set a switching window, the two
buttons 

SP

and 

must be pressed

simultaneously before the power supply is
switched on.

Release these buttons again after the power
supply has been switched on and the display
test has run. The display will then show the
operating mode.

The cursor buttons 

,

can now be used to

change the operating mode until "FEn"
(standing for "Window") in the display. When
this has been done, press 

SP

to store the

setting. The button 

SP

can be used to display

the switching-point setting, which can be
modified by means of the cursor buttons 

.

The distance between the switching point and
reset point is the switching window. If the
switching point is lower than the reset point, a
signal will be output as long as the pressure 
lies within the preset window (case 1, rising
pressure). If the switching point is higher than
the reset point, a signal will be output as long 
as the pressure lies outside the preset window
(case 2, rising pressure). In the case of falling
pressure, the signal is inverted.

Case 1

Case 2

Std = Standard mode, switching and reset 

points adjustable

HYS = Hysteresis mode, switching point and 

hysteresis adjustable

FEn = Window mode, switching window 

adjustable

Q

Q

p

p

Q

Q

p

p

I

N

S

T

R

U

C

T

I

O

N

 

M

A

N

U

A

L

Off = Pressure within window

1

0

RP

<

SP

p

On = Pressure within window

SP

<

RP

p

1

0

Positive hysteresis

1

0

Switching point

p

Negative hysteresis

1

0

Switching point

p

A

DJUSTING THE

S

WITCHING

P

OINTS

(SP) 

AND

R

ESET

P

OINTS

(RP)

Содержание PSW32 SERIES

Страница 1: ...User s Guide www omega com e mail info omega com PSW32 SERIES ELECTRONIC PRESSURE SWITCH FOR HYDRAULIC APPLICATIONS ...

Страница 2: ...p 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190 Trappes Tel 33 0 130 621 400 FAX 33 0 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 0 7056 3017 FAX 49 0 7056 8540 Toll Free in Germany 0800 TC OMEGA e mail germany omega com United Kingdom One Omega Drive River Bend Technology Centre ISO...

Страница 3: ...100 of FS Value Linearity 0 5 of FS 1 digit Electrical Connector Without Analog Output DIN 43650 Table A with Removable Cable Plug Adapter With Analog Output DIN 43651 7 Pin Circular Connector FEATURES Real time LED Status Display of Pressure Adjustable Hysteresis Off line Calibration Fast Accurate Response Extensive Service Life Analog Output Optional APPLICATION IIdeal for Control of Air Water G...

Страница 4: ... RP Load 1 24V 3 Uout Imax 1A 2 0V PE I Micro Controller SP RP p E2 PROM SP RP Load 1 73 44 1 57 40 1 57 40 47 12 59 15 A A Ø 12 3 94 24 1 4 18 NPT 22 5 5 22 5 5 View on A M5 35 9 45 11 5 79 20 Mounting Plates Included 1 4 NPT Pressure Port 304 Stainless Steel All dimensions in inches mm PART NUMBER IDENTIFICATION ELECTRICAL PARAMETERS Electrical connection DIN 43650 Table A without analog output ...

Страница 5: ...ad of the reset point this value can be selected as desired In order to set a hysteresis value the two buttons SP and must be pressed simultaneously before the power supply is switched on Release these buttons again after the power supply has been switched on and the display test has run The display will then show the operating mode The cursor buttons can now be used to change the operating mode u...

Страница 6: ...d power supply Check wiring If the switching line from the load e g electrical control device PLC or similar is being held at an open circuit potential of 3V or if several pressure switches are being operated in parallel this function should be switched off Disconnection during display test then adjust with or U Lo Voltage low Power supply voltage too low 17V Check power supply Load may be too lar...

Страница 7: ...wise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequen tial incidental or special damages CONDITIONS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical appli cations or used on humans Sh...

Страница 8: ...IVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Data Acquisition Engineering Software M U Communications Based Acquisition Systems M U Plug in Cards for Apple IBM Compatibles M U Recorders Printers Plotters HEATERS M U Heating Cable Cartridge Strip Heaters M U Immer...