Configuration Aspects for Programmers
Appendix C
C-12
NetScan User’s Manual
To enable this functionality you must perform the following steps:
1.
Set the first parameter of the Counts (
Y
) command to -1. If more than one transition cycle is
desired, enable the Auto-Rearm flag of the Trigger Configuration command. If this is done, each
transition cycle will be stored in its own Trigger Block within the buffer.
2.
Set the timebase intervals for each event using the Timebase Interval (
I
) command.
3.
Specify the two trigger events via the Trigger Configuration (
T
) command.
The following graph shows how the above application would be organized within the buffer. Using this method,
continuous data may be collected as long as the application program can read data out of the buffer before the
buffer overruns. If the buffer overruns, the data read may not be continuous.
Содержание OMB-NETSCAN
Страница 6: ...iv NetScan User s Manual...
Страница 18: ...1 12 Configuring and Starting NetScan NetScan User s Manual Notes...
Страница 38: ...3 8 General Information and Specifications NetScan User s Manual Notes...
Страница 82: ...4 44 ChartView Software Reference NetScan User s Manual Notes...
Страница 111: ...NetScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 118: ...6 20 Calibration NetScan User s Manual...
Страница 140: ...A ii NetScan User s Manual...
Страница 192: ...API Command Reference Appendix A A 52 NetScan User s Manual Notes...
Страница 237: ...Appendix D Registers Data Formats Queries NetScan User s Manual D 13...
Страница 244: ...NetScan Program Examples Appendix E E 2 NetScan User s Manual...
Страница 248: ...ASCII Code Summary Appendix F F 4 NetScan User s Manual Notes...
Страница 250: ...NetScan Error Messages Appendix G G 2 NetScan User s Manual Notes...
Страница 252: ...Abbreviations Appendix H H 2 NetScan User s Manual Notes...
Страница 254: ...NetScan User s Manual...