7-12
NetScan User’s Manual
CSN14/TC/P
Thermocouple and Low Volts Card with Subminiature Plugs
The CSN14/TC/P card contains 16 differential input channels, each of which may be configured as any
thermocouple type or as a milli-volt input. Temperature values may be returned in units of °C, °F, °K, °R, or mV.
This signal conditioning card contains “subminiature plug type” connectors. There are no user configuration switches
or jumpers on the CSN14/TC/P card.
Number of Channels:
•
16 differential; programmable by channel for specific thermocouple type or voltage.
Input Types:
•
J, K, T, E, R, S, B, N; and ±10V, ±5V, ±1V, ±100mV
Input Connector:
•
Subminiature-plug
Temperature Range, Accuracy, and Resolution
1
•
Type J:
-100
°
to +760
°
C
-200
°
to -100
°
C
±
0.5
°
C
± 0.8°C
0.10
°
C
0.20°C
•
Type K:
-100
°
to +1372
°
C
-200°C to -100°C
±
0.6
°
C
± 0.8°C
0.10
°
C
0.20°C
•
Type T:
-100
°
to +400
°
C
-200
°
to -100
°
C
±
0.5
°
C
± 0.8°C
0.15
°
C
0.25°C
•
Type E:
-100
°
to +1000
°
C
-200°C to -100°C
±
0.7
°
C
±
0.9
°
C
0.10
°
C
0.20°C
•
Type R:
0.0
°
to +1780
°
C
±
2.0
°
C
0.40
°
C
•
Type S:
0.0
°
to +1780
°
C
±
2.0
°
C
0.40
°
C
•
Type B:
+350
°
to +1820
°
C
±
2.0
°
C
0.50
°
C
•
Type N:
-100
°
to +1300
°
C
-200° to -100°C
±
0.6
°
C
± 0.9°C
0.15
°
C
0.20°C
Temperature Units:
•
°
C,
°
F,
°
K,
°
R, mV and volts.
Fault Detection:
•
Open T/C may be detected by a software query.
Cold Junction Sensors:
•
One for every 4 input channels.
Input Impedance and Input Bias Current:
•
1M Ohm typical; and 20 nA maximum
Maximum Common Mode Voltage
4
:
•
200 VAC, or 200 VDC peak
Maximum Normal Mode Voltage:
•
10 VDC or 10VAC peak
Common Mode Rejection:
•
100 dB typical
Maximum Allowable Input:
•
±
25V rms
Channel-to-Power Ground
Isolation:
•
± 200V peak
Channel-to-Channel Isolation:
•
±
200V peak.
Temperature Coefficient:
•
< (0.1 x rated accuracy)%/°C
Содержание OMB-NETSCAN
Страница 6: ...iv NetScan User s Manual...
Страница 18: ...1 12 Configuring and Starting NetScan NetScan User s Manual Notes...
Страница 38: ...3 8 General Information and Specifications NetScan User s Manual Notes...
Страница 82: ...4 44 ChartView Software Reference NetScan User s Manual Notes...
Страница 111: ...NetScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 118: ...6 20 Calibration NetScan User s Manual...
Страница 140: ...A ii NetScan User s Manual...
Страница 192: ...API Command Reference Appendix A A 52 NetScan User s Manual Notes...
Страница 237: ...Appendix D Registers Data Formats Queries NetScan User s Manual D 13...
Страница 244: ...NetScan Program Examples Appendix E E 2 NetScan User s Manual...
Страница 248: ...ASCII Code Summary Appendix F F 4 NetScan User s Manual Notes...
Страница 250: ...NetScan Error Messages Appendix G G 2 NetScan User s Manual Notes...
Страница 252: ...Abbreviations Appendix H H 2 NetScan User s Manual Notes...
Страница 254: ...NetScan User s Manual...