background image

USING THE RPS-1in the OM-420

Omega Engineering Inc.

  

7

Stamford, CT

BATTERY REPLACEMENT

NOTE: USE CAUTION IN WORKING WITH THE GEL-CELL BATTERIES TO PREVENT SHORTING OF
THE TERMINALS OR WIRING AT ALL TIMES!

Under reasonable operating conditions and temperatures, the two gel-cell batteries should provide several
hundred charge/discharge cycles.  When the batteries will not charge completely and/or they discharge
rapidly, they will need to be replaced.  To remove the batteries, follow these steps

1. TURN THE MAIN POWER SWITCH OFF
2. Unscrew the four thumbscrews from the battery pack cover (On the inside of the enclosure

door).

3. Disconnect the two plastic wiring connectors.
4. Remove the old batteries and insert the new ones.
5. Reconnect the two plastic wiring connectors, Ensure that all wires match colors.
6. Tuck the connectors between the batteries and the edge of the enclosure door
7. Replace cover & install thumbscrews
8. 

CAUTION:  CHECK THESE CONNECTIONS BEFORE TURNING THE MAIN POWER SWITCH ON,
INCORRECT CONNECTION WILL RESULT IN PERMANENT DAMAGE TO THE RPS-1.

TROUBLESHOOTING

SYMPTOM:  OUTPUT VOLTAGE IS LOW

PROBLEM:

  Batteries are discharged

SOLUTION:

 Recharge the batteries

PROBLEM:

  Power Supply is in current limiting mode

SOLUTION:

 Decrease the load current draw by reducing supply voltage (if compatible with sensor

requirements), switching some of the load to the other Power Supply.

PROBLEM:

  More than one OUTPUT VOLTAGE PROGRAMMING SWITCH is ON.

SOLUTION:

 Review the OUTPUT VOLTAGE setting table and correct the switch settings.

PROBLEM:

  Desired power supply output voltage is 12 VDC or greater and the MAIN POWER

SWITCH is set to 12 VDC.

SOLUTION:

  Switch the MAIN POWER SWITCH to 24 VDC.

SYMPTOM:  OUTPUT VOLTAGE IS CORRECT BUT VOLTAGE AT SENSOR IS LOW

PROBLEM:

  Excessive voltage drop in sensor wiring

SOLUTION:

  Check for bad connections in wiring loop

SOLUTION:

  Use larger gage wire  (18 AWG should be sufficient for most applications)

SYMPTOM:  STATE-OF-CHARGE INDICATOR DOES NOT LIGHT

PROBLEM:

  The MAIN POWER SWITCH is OFF

SOLUTION:

 Switch to 12VDC or 24VDC position

Содержание OM-420

Страница 1: ...ange of 10 to 60C A slide switch allows the User to configure the RPS 1 for 12 or 24VDC operation Two User programmable Power Supply outputs are available which can source 7 different regulated voltag...

Страница 2: ...r the 12VDC or 24VDC position CHARGER INPUT Two terminal connections are provided for connection of an external current source for charging the RPS 1 batteries POWER SUPPLY OUTPUTS Two independent Use...

Страница 3: ...ng circuits For these two reasons ventilation around the package will maximize performance as well as minimize risk of potentially hazardous explosions WIRING The OM 420 is factory configured to use O...

Страница 4: ...tion If the voltage is less than the OUTPUT VOLTAGE SWITCH programmed setting the power supply is in current limit mode and the current load must be reduced In many applications this can be simply don...

Страница 5: ...e and mounting kit information NOTE The MAIN POWER SWITCH must be in the 24VDC or 12VDC position to allow charging BOOST CHARGE The BOOST CHARGE momentary push button switch on the RPS 1 provides init...

Страница 6: ...ternal current limiting The RPS 1 can be configured for this by placing the red pin jumper on the PCB in the EXTERNAL position See Figure 4 This RED colored jumper is located on the back of the RPS 1...

Страница 7: ...CHECK THESE CONNECTIONS BEFORE TURNING THE MAIN POWER SWITCH ON INCORRECT CONNECTION WILL RESULT IN PERMANENT DAMAGE TO THE RPS 1 TROUBLESHOOTING SYMPTOM OUTPUT VOLTAGE IS LOW PROBLEM Batteries are d...

Страница 8: ...OFF during charging SOLUTION Switch must be ON PROBLEM Batteries have insufficient charge to power charging start up circuitry SOLUTION Press and hold the BOOST CHARGE button for approximately 10 sec...

Отзывы: