USING THE RPS-1in the OM-420

Omega Engineering Inc.

  

1

Stamford, CT

USING THE RPS-1, RECHARGEABLE POWER SUPPLY

IN THE OM-420

OVERVIEW:

The RPS-1 is a self-contained battery power source for use in conjunction with the OM-320/420 System
Base to power transducers requiring external power (e.g. Strain Gages, Pressure transducers, 4 to 20ma
Loop Transmitters, etc.) over a voltage range from 3.5 to 22 VDC.

The RPS-1 contains two User-configurable, rechargeable gel-cell type batteries and integral recharging
circuitry.  The charging circuitry will accept AC or DC current from Photovoltaic arrays, wind generators,
utility sourced transformers/plug-in power supplies, or other current sources.  The sealed batteries are
capable of operating in any orientation and over a temperature range of -10 to 60C.  A slide switch allows
the User to configure the RPS-1 for 12 or 24VDC operation.

Two User-programmable Power Supply outputs are available which can source 7 different regulated
voltage levels from 3.5VDC to 22VDC.  Outputs are short circuit / over-current protected.

The two Power Supplies can be independently programmed for output voltage as well as type of
operational mode. Two modes are available, Continuous ON or AUTOMATIC Operation.  In AUTOMATIC
Operation, the power supply is under control of an optically isolated low voltage (5VDC, 0.5mA) control
input signal.  This control INPUT interfaces directly to a OM-320/420 Digital Output line which can be
programmed to cycle the RPS-1 power supplies ON and OFF, providing transducer excitation under OM-
320/420 control during a logging session.  This automatic power supply cycling  technique maximizes RPS-
1 battery life. See Figure 1 for a sample Hypernet

Figure 1: RPS-1 Control Using HyperWare Warm-Up

Icon

Содержание OM-420

Страница 1: ...ange of 10 to 60C A slide switch allows the User to configure the RPS 1 for 12 or 24VDC operation Two User programmable Power Supply outputs are available which can source 7 different regulated voltag...

Страница 2: ...r the 12VDC or 24VDC position CHARGER INPUT Two terminal connections are provided for connection of an external current source for charging the RPS 1 batteries POWER SUPPLY OUTPUTS Two independent Use...

Страница 3: ...ng circuits For these two reasons ventilation around the package will maximize performance as well as minimize risk of potentially hazardous explosions WIRING The OM 420 is factory configured to use O...

Страница 4: ...tion If the voltage is less than the OUTPUT VOLTAGE SWITCH programmed setting the power supply is in current limit mode and the current load must be reduced In many applications this can be simply don...

Страница 5: ...e and mounting kit information NOTE The MAIN POWER SWITCH must be in the 24VDC or 12VDC position to allow charging BOOST CHARGE The BOOST CHARGE momentary push button switch on the RPS 1 provides init...

Страница 6: ...ternal current limiting The RPS 1 can be configured for this by placing the red pin jumper on the PCB in the EXTERNAL position See Figure 4 This RED colored jumper is located on the back of the RPS 1...

Страница 7: ...CHECK THESE CONNECTIONS BEFORE TURNING THE MAIN POWER SWITCH ON INCORRECT CONNECTION WILL RESULT IN PERMANENT DAMAGE TO THE RPS 1 TROUBLESHOOTING SYMPTOM OUTPUT VOLTAGE IS LOW PROBLEM Batteries are d...

Страница 8: ...OFF during charging SOLUTION Switch must be ON PROBLEM Batteries have insufficient charge to power charging start up circuitry SOLUTION Press and hold the BOOST CHARGE button for approximately 10 sec...

Отзывы: