background image

1. 

DISPLAY

: 5 digit LCD 0.48" (12.2 mm) high digits

DP63000A-I

: Reflective LCD with full viewing angle

DP63000B-I

: Transmissive LCD with selectable red or green LED backlight,

viewing angle optimized. Display color change capability with output state
when using an output module.

2. 

POWER

: Input voltage range is +9 to +28 VDC with short circuit and input

polarity protection. Must use a DP6-MLPS1 or a Class 2 or SELV rated power
supply.

3. 

INPUT RANGES

: Jumper Selectable

D.C. Currents

: 200 

µ

A, 2 mA, 20 mA, or 200 mA

4. 

SIGNAL INPUTS:

5. 

OVERRANGE RATINGS, PROTECTION & INDICATION:

9 to 28 VDC power circuit is not isolated from the signal circuit.

Input Overrange Indication

:  “

OLOL

”.

Input Underrange Indication

:  “

ULUL

”.

Display Overrange/Underrange Indication

: “

.....

”/“-

.....

” 

6. 

DISPLAY RESPONSE TIME:

500 msec min.

7. 

NORMAL MODE REJECTION:

60 dB 50/60 Hz

8. 

USER INPUT (USR):

Programmable input. Connect terminal to common

(USR COMM) to activate function. Internal 10K

pull-up resistor to +9 to

28 VDC.

Threshold Levels

: V

IL

= 1.0 V max; V

IH

= 2.4 V min; V

MAX

= 28 VDC 

Response Time

: 5 msec typ.; 50 msec debounce (activation and release)

9. 

MEMORY

: Nonvolatile E

2

PROM memory retains all programming

parameters and max/min values when power is removed.

10

ENVIRONMENTAL CONDITIONS

:

Operating Temperature Range for DP63000A-I

: -35 to 75°C

Operating Temperature Range for DP63000B-I depends on display color

and intensity level as per below

:

Storage Temperature

: -35 to 85°C

Operating and Storage Humidity

: 0 to 85% max. relative humidity (non-

condensing)

Vibration According to IEC 68-2-6

: 5 to 500 Hz, in X, Y, Z direction for 1.5

hours, 5 g’s.

Shock According to IEC 68-2-27

: Operational 30 g, 11 msec in 3 directions

Altitude

: Up to 2000 meters

11. 

CONNECTIONS

: Wire clamping screw terminals

Wire Strip Length

: 0.3" (7.5 mm)

Wire Gage

: 30-14 AWG copper wire

Torque

: 5 inch-lbs (0.565 N-m) max.

12. 

CONSTRUCTION

: This unit is rated for NEMA 4X/IP65 requirements for

indoor use. Installation Category I, Pollution Degree 2. High impact plastic
case with clear viewing window. Panel gasket and mounting clip included.

13.

CERTIFICATIONS AND COMPLIANCES

:

SAFETY

UL Recognized Component, File #E313607, UL61010A-1, CSA 22.2 No. 61010-1 

Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.

UL Listed, File # E313547, UL508, CSA C22.2 No. 14-M95

LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment

for measurement, control, and laboratory use, Part 1.

IP65 Enclosure rating (Face only), IEC 529
Type 4X Indoor Enclosure rating (Face only), UL50

ELECTROMAGNETIC COMPATIBILITY

Emissions and Immunity to EN 61326: Electrical Equipment for Measurement,

Control and Laboratory use.

Notes:
1. Criterion A: Normal operation within specified limits.

14. 

WEIGHT

: 3.2 oz (100 g)

4

G

ENERAL

M

ETER

S

PECIFICATIONS

MODEL

NO.

DISPLAY COLOR

INPUT CURRENT

@ 9 VDC WITH 

DP6-RLY0

DP63000A-I

---

10 mA

40 mA

DP63000B-I

Red (max intensity)

85 mA

115 mA

DP63000B-I

Green (max intensity)

95 mA

125 mA

TEMPERATURE

1 & 2

-35 to 75°C

3

-35 to 70°C

4

-35 to 60°C

5

-35 to 50°C

1 & 2

-35 to 75°C

3

-35 to 65°C

5

4

-35 to 35°C

-35 to 50°C

INTENSITY LEVEL

Class A

EN 55011

Emissions

Emissions:

30 A/m 

3 V/rms 
Criterion A

Criterion A

EN 61000-4-8

EN 61000-4-6

RF conducted interference

1 kV L-L, 

Criterion A

EN 61000-4-5

Surge 

1 kV signal

2 kV power

Criterion A

EN 61000-4-4

Fast transients (burst)

2 kV L&N-E power

10 V/m

Criterion A

EN 61000-4-3

Electromagnetic RF fields

8 kV air discharge

4 kV contact discharge

Criterion A

EN 61000-4-2

Electrostatic discharge

Immunity to Industrial Locations:

Power frequency magnetic fields

Green Display

Red Display

INPUT CURRENT

@ 9 VDC WITHOUT

DP6-RLY0

INPUT 

RANGE

ACCURACY

@23 °C, less 

than 85% RH

INPUT 

IMPEDANCE

MAX 

INPUT 

SIGNAL

RESOLUTION

TEMP.

COEFFICIENT

200 

µ

A

0.1% of span

1.111 K

15 mA

10 nA

70 ppm / °C

2 mA

0.1% of span

50 mA

.1 µA

70 ppm / °C

20 mA

0.1% of span

11 

150 mA

1 µA

70 ppm / °C

200 mA 0.1% of span

500 mA

10 µA

70 ppm / °C

111 

P

ART

N

UMBER

I

NFORMATION

DP6-SNK0

Dual Sinking Open Collector Output card

DP6-SOFT

DP6-485-CABLE

RS485 Programming Cable (DB9-RJ11)

DP63000B-I

PC Configuration Software for Windows 98, ME, 2000, XP

*

DC Current Meter with backlight display

DP6-232-CABLE

DP6-RLY0

DP6-MLPS1

Micro-Line Power Supply, 85 to 250 VAC

Single Relay Output Card

RS232 Programming Cable (DB9-RJ11)

DP6-COM2

RS232 Serial Communications Card

DP6-COM1

RS485 Serial Communications Card

DP63000A-I

DC Current Meter with reflective display

PART NUMBER

DESCRIPTION

Содержание DP63000A-I

Страница 1: ...LP0694A omega com e mail info omega com For latest product manuals omegamanual info User sGuide Shop online at D DP P6 63 30 00 00 0A A I I D DP P6 63 30 00 00 0B B I I Digital P Panel M Meter...

Страница 2: ...megashop cz Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega D...

Страница 3: ...s Serial communications capability for RS232 or RS485 is added with a serial option module The DP63000 can be powered from an optional Power Supply Model Number DP6 MLPS1 that attaches directly to the...

Страница 4: ...1 Recognized to U S and Canadian requirements under the Component Recognition Program of Underwriters Laboratories Inc UL Listed File E313547 UL508 CSA C22 2 No 14 M95 LISTED by Und Lab Inc to U S and...

Страница 5: ...ver the rear of the unit to the back of the bezel The unit should be installed fully assembled Insert the unit into the panel cutout While holding the unit in place push the panel latch over the rear...

Страница 6: ...commercial radio transmitter 4 Signal or Control cables within an enclosure should be routed as far as possible from contactors control relays transformers and other noisy components 5 In extremely h...

Страница 7: ...tions must then be given to the potential of the user input common with respect to earth ground and the common of the plug in cards with respect to input common Before connecting signal wires the Inpu...

Страница 8: ...he value and move to the next parameter PROGRAMMING MODE EXIT SEL BUTTON The Programming Mode is exited by pressing the SEL button with Pro NO displayed This will commit any stored parameter changes t...

Страница 9: ...KEY style enter the first Input Value using the front panel buttons The Input Range selection sets the decimal location for the Input Value For Apply APLY style the meter shows the previously stored I...

Страница 10: ...current calibration values to provide accurate current measurements Over time the electrical characteristics of the components inside the unit will slowly change with the result that the stored calib...

Страница 11: ...n allows the SEL button to toggle through the enabled displays FRONT PANEL DISPLAY SELECT ENABLE SEL This selection allows the RST button to reset the selected value s FRONT PANEL RESET ENABLE RST The...

Страница 12: ...en the setpoint is a control output usually balanced hysteresis is used For alarm applications usually unbalanced hysteresis is used For unbalanced hysteresis modes the hysteresis functions on the low...

Страница 13: ...r input or RST button is activated momentary action the meter delays the event until the corresponding on output crosses the trigger off point Previously latched outputs are off if power up Display Va...

Страница 14: ...PROGRAMMING QUICK OVERVIEW Press and hold SEL button to enter Programming Mode...

Страница 15: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 16: ...rs Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conducti...

Отзывы: