background image

5

ADDING OPTION CARDS

The DP63000 meters can be fitted with optional output cards and/or serial

communications cards. The details for the plug-in cards can be reviewed in the
specification section below. The plug-in cards, that are sold separately, can be
installed initially or at a later date.

WARNING: Disconnect all power to the unit before

installing Plug-in card.

Note

: Measurement errors may occur if signal input common is shared with

another circuit common (ie, serial common, Dual Sinking Output option card,
or Power Supply common) on multiple units.

SINGLE RELAY CARD

Type

: Single FORM-C relay

Isolation To Sensor & User Input Commons

: 1400 Vrms for 1 min.

Working Voltage: 150 Vrms

Contact Rating

: 1 amp @ 30 VDC resistive; 0.3 amp @ 125 VAC resistive

Life Expectancy

: 100,000 minimum operations

Response Time

:

Turn On Time: 4 msec max.
Turn Off Time: 4 msec max.

DUAL SINKING OUTPUT CARD

Type

: Non-isolated switched DC, N Channel open drain MOSFET

Current Rating

: 100 mA max. 

V

DS ON

: 0.7 V @ 100 mA

V

DS MAX

: 30 VDC

Offstate Leakage Current

: 0.5 mA max.

RS485 SERIAL COMMUNICATIONS CARD

Type

: RS485 multi-point balanced interface (non-isolated)

Baud Rate

: 300 to 38.4k

Data Format

: 7/8 bits; odd, even, or no parity

Bus Address

: 0 to 99; max 32 meters per line

Transmit Delay

: Selectable (refer to DP6-COM bulletin)

RS232 SERIAL COMMUNICATIONS CARD

Type

: RS232 half duplex (non-isolated)

Baud Rate

: 300 to 38.4k

Data Format

: 7/8 bits; odd, even, or no parity

O

PTIONAL

P

LUG

-

IN

C

ARDS

1.0 I

NSTALLING THE

M

ETER

INSTALLATION 

The meter meets NEMA 4X/IP65 requirements when properly installed. The

unit is intended to be mounted into an enclosed panel. Prepare the panel cutout
to the dimensions shown. Remove the panel latch from the unit. Slide the panel
gasket over the rear of the unit to the back of the bezel. The unit should be
installed fully assembled. Insert the unit into the panel cutout. 

While holding the unit in place, push the panel latch over the rear of the unit

so that the tabs of the panel latch engage in the slots on the case. The panel latch
should be engaged in the farthest forward slot possible. To achieve a proper seal,
tighten the latch screws evenly until the unit is snug in the panel (Torque to
approx. 28 to 36 in-oz [0.202 to 0.26 N-m]). Do not over-tighten the screws.

INSTALLATION ENVIRONMENT

The unit should be installed in a location that does not exceed the operating

temperature and provides good air circulation. Placing the unit near devices that
generate excessive heat should be avoided.

The bezel should only be cleaned with a soft cloth and neutral soap product.

Do NOT use solvents. Continuous exposure to direct sunlight may accelerate the
aging process of the bezel.

Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate the

keypad of the unit.

2.0 S

ETTING THE

J

UMPERS

INPUT RANGE JUMPER

This jumper is used to select the proper input range. The input range selected

in programming must match the jumper setting. Select a range that is high
enough to accommodate the maximum signal input to avoid overloads. To
access the jumper, remove the rear cover of the meter. 

Warning

: Exposed line voltage exists on the circuit boards. Remove

all power to the meter and load circuits before accessing inside of
the meter.

REMOVING THE REAR COVER

To remove the rear cover, locate the cover locking tab below the 2nd and 3rd

input terminals. To release the tab, insert a small, flat blade screwdriver between
the tab and the plastic wall below the terminals. Inserting the screwdriver will
provide enough pressure to release the tab locks. To replace the cover, align the
cover with the input terminals and press down until the cover snaps into place.

Содержание DP63000A-I

Страница 1: ...LP0694A omega com e mail info omega com For latest product manuals omegamanual info User sGuide Shop online at D DP P6 63 30 00 00 0A A I I D DP P6 63 30 00 00 0B B I I Digital P Panel M Meter...

Страница 2: ...megashop cz Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 9398 0 FAX 49 0 7056 9398 29 Toll Free in Germany 0800 639 7678 e mail info omega de United Kingdom One Omega D...

Страница 3: ...s Serial communications capability for RS232 or RS485 is added with a serial option module The DP63000 can be powered from an optional Power Supply Model Number DP6 MLPS1 that attaches directly to the...

Страница 4: ...1 Recognized to U S and Canadian requirements under the Component Recognition Program of Underwriters Laboratories Inc UL Listed File E313547 UL508 CSA C22 2 No 14 M95 LISTED by Und Lab Inc to U S and...

Страница 5: ...ver the rear of the unit to the back of the bezel The unit should be installed fully assembled Insert the unit into the panel cutout While holding the unit in place push the panel latch over the rear...

Страница 6: ...commercial radio transmitter 4 Signal or Control cables within an enclosure should be routed as far as possible from contactors control relays transformers and other noisy components 5 In extremely h...

Страница 7: ...tions must then be given to the potential of the user input common with respect to earth ground and the common of the plug in cards with respect to input common Before connecting signal wires the Inpu...

Страница 8: ...he value and move to the next parameter PROGRAMMING MODE EXIT SEL BUTTON The Programming Mode is exited by pressing the SEL button with Pro NO displayed This will commit any stored parameter changes t...

Страница 9: ...KEY style enter the first Input Value using the front panel buttons The Input Range selection sets the decimal location for the Input Value For Apply APLY style the meter shows the previously stored I...

Страница 10: ...current calibration values to provide accurate current measurements Over time the electrical characteristics of the components inside the unit will slowly change with the result that the stored calib...

Страница 11: ...n allows the SEL button to toggle through the enabled displays FRONT PANEL DISPLAY SELECT ENABLE SEL This selection allows the RST button to reset the selected value s FRONT PANEL RESET ENABLE RST The...

Страница 12: ...en the setpoint is a control output usually balanced hysteresis is used For alarm applications usually unbalanced hysteresis is used For unbalanced hysteresis modes the hysteresis functions on the low...

Страница 13: ...r input or RST button is activated momentary action the meter delays the event until the corresponding on output crosses the trigger off point Previously latched outputs are off if power up Display Va...

Страница 14: ...PROGRAMMING QUICK OVERVIEW Press and hold SEL button to enter Programming Mode...

Страница 15: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 16: ...rs Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conducti...

Отзывы: