  After observing the specimen with the 100X objective, clean the 100X objective 

lens first.

 

3.7  Adjusting condenser 

1)  Turn the condenser focus knob to raise or lower the condenser.   
2)  Raise the condenser when using high power objectives and lower it when using 

low power objectives. 

Note:   

 

The centering of the condenser and the light axis of the objective are factory 
adjusted. Do not attempt re-adjustment. 

 

 

The highest position of the condenser has been factory adjusted. Do not attempt 
re-adjustment.

 

3.8  Adjusting iris aperture diaphragm 

Swing the iris diaphragm lever (

Fig. 7

) left or right to 

adjust the aperture size. 

Note:   

 

The iris diaphragm is designed to adjust the aperture 
size, not to adjust the brightness although the 
brightness will be changed when it's adjusted. When 
aperture is adjusted to smaller size, the contrast will be 
increased and the depth of field will be increased as 
well. Turn up the intensity of the light if the image is too dim

.

 

3.9  Camera 

(optional, may not included in your package)

 

1) Bring the microscope into focus by following the procedures in 

3.5

.   

2) Install the camera by following the procedures in 

2.4

.   

3) Open image observing software to examine. 
4) You also can capture images or record live videos through the software, depending 

on the functions provided by the software. 

Note:

   

Please refer to the manuals in the c

amera’s CD for the details of installation and 

operation of the camera.

 

3.10  Dry Darkfield condensers observation

 (optional, may not included in your 

package)

 

See section 7 

(

Dry Darkfield Condenser Installation & Operation Instruction

 

 
 
 
 
 
 
 
 
 
 

Fig. 7 

Iris Diaphragm Lever 

Содержание M82ES Series

Страница 1: ...User Manual Binocular Compound Microscope Model M82ES Series MicroscopeNet com...

Страница 2: ...aution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 8 5 Optional Parts 9 6 Troubleshooting Guide 10 7 Dry Darkfield Condenser Installation Ope...

Страница 3: ...t of direct sunlight high temperature or humidity and dusty environments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto th...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...5 Objective 18 Power Switch 6 Slide Holder 19 Brightness Intensity Dial 7 Mechanical Stage 20 Immersion Oil 8 Light Collector 21 Dust Cover 9 Microscope Base 22 Condenser Lock Thumb Screw 10 Eyepiece...

Страница 6: ...lled securely 2 2 Installation of the eyepieces 1 Remove the protective caps from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 3 Installation of the glass filter 1 Swing out the...

Страница 7: ...ve to touch a specimen slide when changing objectives 3 3 Adjusting interpupillary distance While observing with both eyes hold the left and right eyepiece tubes and slide the tubes in and out The int...

Страница 8: ...iece to seat the observing position between the 40X and 100X objectives as shown in Fig 6 a 2 Place a drop of immersion oil on the slide cover as shown in Fig 6 b 3 Rotate the objective nosepiece to s...

Страница 9: ...adjust the brightness although the brightness will be changed when it s adjusted When aperture is adjusted to smaller size the contrast will be increased and the depth of field will be increased as we...

Страница 10: ...0 17 spring oil Condenser Abbe NA 1 25 w iris diaphragm Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double layer di...

Страница 11: ...200X 400X 800X 1000X 2000X 5 Optional Parts The optional parts may be included in some models or sold separately 1 Darkfield Condensers Model Darkfield Condenser Numerical Aperture Objective Mounting...

Страница 12: ...sed image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at t...

Страница 13: ...e the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn the...

Отзывы: