Troubleshooting Guide 

 

Problem

 

Cause

 

Solution

 

Lamp does not light when 
switched on 

No electrical power   

Check power cord connection   

LED or power unit dead 

Contact seller for service   

Darkness at the periphery 

Revolving nosepiece not in click stop 
position 

Revolve the nosepiece to click-stop 
position by swinging the objective correctly 
into the optical path 

Uneven brightness in the 
field of view 

The diffusion filter is not in the light path 

Check the diffusion filter and make sure it’s 
in the light path 

Insufficient brightness 

Diffusion filter used when using 100X 
objective 

Swing out the diffusion filter when using 
100X objective 

Dirt or dust on the view 

Dirt or dust on the lens eyepiece, 
condenser, objective, collector lens or 
specimen 

Clean the lens with a camera cleaning kit 

Dirt or dust on the specimen 

Clean the specimen 

Poor image quality or not 
able to get focused image 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down (specimen at 
the bottom) 

Turn slide over so the cover-glass faces up 

Diopter adjustment is not set properly 

Readjust the diopter settings 

Immersion oil is on a dry objective 
(especially the 40X) 

Check the objectives, clean if necessary 

No immersion oil used with 100X 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Specimen rises from stage surface 

Secure the specimen in the slide holder 

Condenser aperture is closed or open 
too much 

Open or close properly 

Condenser is positioned too low or too 
high 

Move condenser upper or lower 

Lamp intensity is too high or low 

Adjust the light intensity by rotating the 
intensity control dial 

Image is black screen 
through camera 

(

optional)

 

No enough lighting or driver not installed 
properly 

Check the camera manual 

Image is totally white 
through camera 

(

optional)

 

The light is too bright 

Manually adjust the exposure in the 
software. See manual for details. 

 

Содержание M82ES Series

Страница 1: ...User Manual Binocular Compound Microscope Model M82ES Series MicroscopeNet com...

Страница 2: ...aution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 8 5 Optional Parts 9 6 Troubleshooting Guide 10 7 Dry Darkfield Condenser Installation Ope...

Страница 3: ...t of direct sunlight high temperature or humidity and dusty environments Ensure the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto th...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...5 Objective 18 Power Switch 6 Slide Holder 19 Brightness Intensity Dial 7 Mechanical Stage 20 Immersion Oil 8 Light Collector 21 Dust Cover 9 Microscope Base 22 Condenser Lock Thumb Screw 10 Eyepiece...

Страница 6: ...lled securely 2 2 Installation of the eyepieces 1 Remove the protective caps from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 3 Installation of the glass filter 1 Swing out the...

Страница 7: ...ve to touch a specimen slide when changing objectives 3 3 Adjusting interpupillary distance While observing with both eyes hold the left and right eyepiece tubes and slide the tubes in and out The int...

Страница 8: ...iece to seat the observing position between the 40X and 100X objectives as shown in Fig 6 a 2 Place a drop of immersion oil on the slide cover as shown in Fig 6 b 3 Rotate the objective nosepiece to s...

Страница 9: ...adjust the brightness although the brightness will be changed when it s adjusted When aperture is adjusted to smaller size the contrast will be increased and the depth of field will be increased as we...

Страница 10: ...0 17 spring oil Condenser Abbe NA 1 25 w iris diaphragm Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with rack stop Mechanical Stage Double layer di...

Страница 11: ...200X 400X 800X 1000X 2000X 5 Optional Parts The optional parts may be included in some models or sold separately 1 Darkfield Condensers Model Darkfield Condenser Numerical Aperture Objective Mounting...

Страница 12: ...sed image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at t...

Страница 13: ...e the bolt to lock it when needed Note When raising the mechanical stage do not make contact with the objective 2 Centering the dry darkfield condenser Turn the 4X objective to the light path Turn the...

Отзывы: