Troubleshooting Guide 

 

OPTICAL PROBLEMS

 

Problem

 

Cause

Solution

 

Darkness at the 
periphery or uneven 
brightness in the 
field of view 

Revolving nosepiece not in click 
stop position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

The light source of the bulb is not at 
the center 

Adjust the position of the bulb 

Dirt or dust on the 
view 

Dirt or dust on the lens eyepiece, 
condenser, objective, collector lens 
or specimen 

Clean the lens with a camera 
cleaning kit 

Poor image quality 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down 

Turn slide over so the cover-glass 
faces up 

Immersion oil is on a dry objective 
(especially the 40x) 

Check the objectives, clean if 
necessary 

No immersion oil used with 100x 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture is closed or 
open too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

 

 

ELECTRICAL PROBLEMS 

Problem

 

Cause

 

Solution

 

Lamp does not light 
when switched on 

No electrical power   

Check power cord connection   

Lamp bulb burnt out   

Replace bulb   

Fuse blown out 

Replace fuse 

 

 

 

 

 

 

 

Содержание M828T

Страница 1: ...User Manual Binocular Compound Microscope Model M828T MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 3 ii Care and Maintenance 4 1 Component Illustration 5 2 Installation 6 3 Operation 8 4 Specifications 10 5 Troubleshooting Guide 11...

Страница 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...tch 3 Eyepiece Tube 11 Head Thumb Screw 19 Condenser Focus Knob 4 Nosepiece 12 Microscope Body 20 Focus Stop Lever 5 Objective 13 Focus Tension Ring 21 Abbe Condenser 6 Slide Holder 14 Coarse Focus Kn...

Страница 6: ...hen changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical light path 2 4 Installation of filter 1 Swing...

Страница 7: ...el may not have one 1 Unplug the power cord 2 Screw off the light cover on the base 3 Screw the black disc onto the base and then insert the mirror into the hole at the center of the black disc You ma...

Страница 8: ...e is the same as your interpupillary distance for example 65 in the right figure 3 Close your left eye and bring the specimen into focus following the focusing procedures in 3 5 4 Close your right eye...

Страница 9: ...diaphragm is designed to adjust the aperture size not to adjust brightness Generally opening the diaphragm to 70 80 of the N A value of the respective objective will provide an image of acceptable qua...

Страница 10: ...ocusing adjustment at 0 002mm range 28mm Stage Mechanical stage Dimension 5 1 2 x 5 1 2 140mmx140mm Movement range 3 x 2 75mm X 50mm Illumination Transmitted 6V 20W Halogen Variable intensity Power Su...

Страница 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Страница 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...

Отзывы: