Installation  

2.1  Installation of the binocular viewing head 

1)  Loosen the thumb screw on the head of the body and remove the plastic cover on 

the top   

2)  Remove the cap on the dovetail of the binocular viewing head 
3)  Insert the binocular viewing head into the head of the body; ensure that the dovetail 

is completely seated into the socket; tighten the thumb screw.   

Caution

: Do not release the head from your hand grip until you are sure the head is 

installed securely.   

2.2  Installation of the eyepieces   

1)  Remove the protective caps from the eyepiece tubes.   
2)  Insert the eyepieces into the eyepiece tubes.     

2.3  Installation of the objectives 

1)  Adjust the coarse focus knob until the mechanical stage is at its lowest position. 
2)  Install the 4x objective into the nosepiece. Then in a clock-wise direction, rotate the 

nosepiece and install each succeeding higher magnification objective. 

Note: 

Inspect the objectives frequently for dirt or oil; clean if necessary. Use the 10x 

objective to initially focus the image of your specimen. When changing the objective 
magnification, rotate the objective nosepiece until you hear a “click” sound. This 
ensures the objective is centered in the optical light path.

 

2.4  Installation of filter 

1)  Swing out the filter holder under the condenser.   
2)  Insert the filter into the holder, swing the holder in.   

2.5  Installing (or changing) the halogen bulb 

   

 

1)  Turn the power off and disconnect the power cord.   
2)  Allow some time to cool down the lamp.   
3)  Turn over the microscope on its side; find the bulb compartment at 

the bottom. 

4)  Open the cover of the bulb compartment by loosening the thumb 

screw. Take out the dead bulb and insert the new bulb. Be sure 
the pins on the bulb are completely inserted into the lamp socket. 
You may also loosen the two screws on the cover to adjust the 
position of the bulb to get centered and even brightness. Screw the cover on.   

Caution: 

Before you turn over the microscope, be sure to take the eyepieces off and be 

certain that the head is securely locked by the thumb screw. 

2.6  Replacing the fuse 

1)  Turn off the power and disconnect the power cord.   
2)  Turn over the microscope on its side; find the fuse at the bottom of the base. 
3)  Turn the fuse holder counter-clockwise to take it off, insert new fuse, and then turn it 

on clockwise. 

Содержание M828T

Страница 1: ...User Manual Binocular Compound Microscope Model M828T MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 3 ii Care and Maintenance 4 1 Component Illustration 5 2 Installation 6 3 Operation 8 4 Specifications 10 5 Troubleshooting Guide 11...

Страница 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...tch 3 Eyepiece Tube 11 Head Thumb Screw 19 Condenser Focus Knob 4 Nosepiece 12 Microscope Body 20 Focus Stop Lever 5 Objective 13 Focus Tension Ring 21 Abbe Condenser 6 Slide Holder 14 Coarse Focus Kn...

Страница 6: ...hen changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical light path 2 4 Installation of filter 1 Swing...

Страница 7: ...el may not have one 1 Unplug the power cord 2 Screw off the light cover on the base 3 Screw the black disc onto the base and then insert the mirror into the hole at the center of the black disc You ma...

Страница 8: ...e is the same as your interpupillary distance for example 65 in the right figure 3 Close your left eye and bring the specimen into focus following the focusing procedures in 3 5 4 Close your right eye...

Страница 9: ...diaphragm is designed to adjust the aperture size not to adjust brightness Generally opening the diaphragm to 70 80 of the N A value of the respective objective will provide an image of acceptable qua...

Страница 10: ...ocusing adjustment at 0 002mm range 28mm Stage Mechanical stage Dimension 5 1 2 x 5 1 2 140mmx140mm Movement range 3 x 2 75mm X 50mm Illumination Transmitted 6V 20W Halogen Variable intensity Power Su...

Страница 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Страница 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...

Отзывы: