background image

 

PV9353F4E007 21 

4.10. Acquiring a background value 

(1) 

Click BACKGROUND1 on the operation panel. 
A clicky sound is generated, and then a warning alarm sounds as the process 
finishes. 
As the data is saved, an asterisk (*) appears at the left of BACKGROUND1 in the 
operation panel. 

 

4.11. Acquiring a reference value 

(1) 

Click REFERENCE on the operation panel. 
The shutter sounds, the reference value is captured. After that, the shutter operates 
again, and a warning alarm sounds as the process finishes. 
As the data is saved, an asterisk (*) appears at the left of REFERENCE in the 
operation panel. 

4.12.  Setting a sample to be measured 

(1)  Set the sample on the top surface of the tilt stage. Locate the sample in the hole at the 

center of the stage. 

 

4.13. Adjusting the position of a sample to be measured 

(1) Adjust the tilt of the sample. 

See 4.8 Adjusting the position of the 45-degree reference sample (3), and do the same 
for the sample tilt adjustment. 

 
(2)  Adjust the shift of the sample. 

See 4.8 Adjusting the position of the 45-degree reference sample (4), and do the same 
for the sample shift adjustment. 

 

4.14. Spectrometry 

(1)  Click MEASURE on the operation panel. 

The shutter sounds, indicating the start of spectroscopy. When alarm sounds and 
spectroscopy is ended, a shutter sounds again. 
The result is displayed in the sub-windows. Up to 20 results can be stored. 
 

Содержание USPM-45EXP

Страница 1: ...sure to read this instruction manual before use Since the description of this document is limited to contents related to the 45 degree expansion unit please also read the separate NIR Micro Spectropho...

Страница 2: ...of this Product 6 1 Checking Delivered Items 7 2 Part Names and Functions Specifications 9 3 Installation and Connection 12 4 Operations 14 5 Notes on Storing 24 6 If Abnormality Occurs 24 7 Conformi...

Страница 3: ...aler or our sales department for replacement or other inquiries Marks Meanings Do not touch because the surface is hot Otherwise you can get burned This indicates a warning or caution covering a wide...

Страница 4: ...this product are as follows Indicates the proper upper direction of the packaged cargo Indicates that the packaged cargo must not be exposed to rain Indicates that packaged cargo must be carefully han...

Страница 5: ...question regarding the content of this operation manual please contact your dealer or our sales department Prohibition of repairs or modifications to the product Do not disassemble or modify this pro...

Страница 6: ...g Strictly observe the following when handling this product In addition observe the warnings or cautions described in each chapter Gently operate each adjustment knob Gently turn the adjustment knobs...

Страница 7: ...e product free of charge 2 This guarantee is not applied to any failure that has been caused by another reason for example a failure in a peripheral device or a problem in the installation location en...

Страница 8: ...you should notice any mistake or omission regarding a description please inquire to the contact point listed in the attached sheet Notwithstanding the foregoing we cannot be held responsible for any...

Страница 9: ...r dealer or a branch of our company It is a reference sample of 45 degrees for the reflection measurement Reference data CD is attached Because the mirror court is given I will recommend the update of...

Страница 10: ...included 1 45 degree reflection measurement reference sample collimator unit for transmittance tilt stage package Name Quantity Remarks 5 collimator unit for transmittance See Note 1 1 6 Tilt stage S...

Страница 11: ...ce Light source unit for optional transmission measurement or 45 degree measurement to be used by fixing the lamp house using the fixing knobs two locations For a detailed explanation see 3 Installati...

Страница 12: ...larization unit Adjust the polarization condition between the P polarization S polarization and random polarization by using the polarization switching knob 5 Tilt stage Adjust the two axis tilt again...

Страница 13: ...ree Model name USPM 45EXP Wavelength range 380 1 050 nm Wavelength resolution Approximately 5 nm wavelength indication resolution 1 nm Repeatability 3 0 3 or less 430 1 010 nm 1 0 or less in other ran...

Страница 14: ...se into the original position The lamp connector must be positioned to the right of the device Check the portion surrounded by a circle below and check the screw hole position and the lamp house fixin...

Страница 15: ...not connected the intended safety performance cannot be ensured 5 Protection key Any USB port on the computer For cable setting keep cables away from something sharp Be careful not to apply excessive...

Страница 16: ...onfigure the work setting file See 4 6 Setting up the work configuration file For the reference settings select FILE and select the file in the attached 45 degree reference sample data CD 6 Set the su...

Страница 17: ...tart the program 4 3 Revolving nosepiece settings 1 Configure the revolving nosepiece settings When executing a 45 degree reflection measurement using the Reflectivity Measurement Unit for 45 Degree a...

Страница 18: ...direction is the direction against the stage surface Refer to the figure below Far position Random polarization Intermediate position P polarization Near position S polarization Polarization switching...

Страница 19: ...tion 1 Choose Setup Work on the File menu The work setup window will open 2 Configure the reference settings 1 Click and select FILE 2 Click the portion surrounded by a circle and open the rf file rf...

Страница 20: ...e FOCUS window and the CT window by double clicking the black portion of the LIVE window The FOCUS window or the CT window can be selected from the file menu Live The FOCUS window is displayed when th...

Страница 21: ...ue of X Y shown at the lower left in the FOCUS window becomes the smallest For a more precise measurement it is recommended that you adjust the value of X Y to X 0 1 Y 0 1 Tilt not in the correct posi...

Страница 22: ...ht amount of the reference light flux reference does not exceed the maximum received light amount of the light receiving element Follow the procedures below to adjust the sampling interval 1 Click SMP...

Страница 23: ...ion panel 4 12 Setting a sample to be measured 1 Set the sample on the top surface of the tilt stage Locate the sample in the hole at the center of the stage 4 13 Adjusting the position of a sample to...

Страница 24: ...urement result 3 DIAGRAM window The measurement result of the chromaticity diagram is displayed The chromaticity diagram can show the XY chromaticity diagram and Lab chromaticity diagram By double cli...

Страница 25: ...the lower limit of the spectroscopic quantity defined in the system configuration window 6 LUMINANCE TEXT window The spectroscopic light amounts of the background reference or measurement are displaye...

Страница 26: ...n Manual 5 3 Wavelength offset 6 If Abnormality Occurs Also check NIR Micro Spectrophotometer USPM W B Operation Manual 6 Troubleshooting 7 Conforming to Laws and Regulations Refer to NIR Micro Spectr...

Страница 27: ...Adjust the revolving nosepiece position to the through position the position where the objective lens is not set 4 Set the BK7 reference sample on the tilt stage 5 Adjust the tilt of the two axes to...

Страница 28: ...s reserved The contents of this manual shall not be reproduced or distributed in part or in whole without Olympus s consent The company names and product names referred to in this document may be the...

Страница 29: ...us Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 Tel 82 2 6255 3458 OLYMPUS THAILAND CO LTD 159 Serm mit Tower 9th floor unit 906 910 Sukhumvit21 road North Klongtoey Wattana Bankok 10110 Thai...

Отзывы: