PV9353F4E007 14
4. Operations
This chapter describes the general operations of the unit. For details on the usage, each all
personnel operating the unit should be sufficiently trained from the viewpoint of each
specialized area.
4.1. Measurement flow
(1)
Start the controller supply unit and computer, and then start the software.
-> See
4.2 Start-up
.
(2)
Configure the revolving nosepiece settings.
-> See
4.3 Revolving nosepiece settings
.
(3)
Optical path switching settings
-> See
4.4 Optical path switching settings
.
(4) Polarization condition settings (for customers who purchased the optional
polarization unit)
-> See
4.5 Polarization condition settings
.
(5)
Configure the work setting file.
-> See
4.6 Setting up the work configuration file
.
For the reference settings, select [FILE], and select the file in the attached 45-degree
reference sample data CD.
(6)
Set the supplied 45-degree reference sample.
-> See
4.7 Setting the 45-degree reference sample
.
(7)
Adjust the position of the reference sample.
-> See
4.8 Adjusting the position of the 45-degree reference sample
.
(8) Set
the
sampling
interval.
-> See
4.9 Adjusting the sampling interval
.
(9)
Acquire the background value.
-> See
4.10 Acquiring a background value
.
(10) Acquire the reference value.
-> See
4.11 Acquiring a reference value
.
(11) Set the sample to be measured.
-> See
4.12 Setting a sample to be measured
.
(12) Adjust the position of the sample to be measured.
-> See
4.13 Adjusting the position of a sample to be measured
.
(13) Execute the measurement.
-> See
4.14 Spectrometry
.