Instrument Overview

 

19

On/off switch (

)

This switch is used to turn on and off the TomoScan FOCUS LT.

POWER

 

CORD

 

PROVIDES

 

AC

 (~) 

POWER

 

DISCONNECT

 

This marking on the back panel indicates that the internal AC power supply is not 
disconnected from the AC source when the unit is turned off using the front panel 
on/off switch (

). To disconnect the AC source, first disconnect the power cord from 

the AC outlet, and then disconnect the power cord from the back of the instrument. 
Failure to do so can cause electric shocks.

Handles

Two handles are located on each side of the front panel. They are used to disassemble 
and reassemble the unit.

The front panel handles should not be used to carry the TomoScan FOCUS LT. 
Because the handles were not designed to transport the instrument, they may break 
off.

Front panel protective bumper

The rubber bumper shown in Figure 1-2 on page 20 protects the front panel 
components.

Содержание TomoScan FOCUS LT

Страница 1: ...ument User s Manual This instruction manual contains essential information on how to use this Olympus product safely and effectively Before using this product thoroughly review this instruction manual...

Страница 2: ...the accuracy of the information contained therein and corresponds to the version of the product manufactured prior to the date appearing on the title page There could however be some differences betw...

Страница 3: ...ruction Manual 5 Instrument Compatibility 5 Repair and Modification 6 Safety Symbols 6 Safety Signal Words 7 Note Signal Words 7 Safety 8 Warnings 8 Battery Precautions 9 Equipment Disposal 10 WEEE Di...

Страница 4: ...Fan Filters 36 4 3 1 Cleaning the Fan Filters 37 4 3 2 Changing the Fan Filters 37 4 4 Fuse Changing 38 4 5 Troubleshooting 40 5 Specifications 41 5 1 General Specifications 41 5 2 Operating Specific...

Страница 5: ...TA041 01EN U8778479 Rev C May 2014 Table of Contents v 6 3 DC IN Connector 63 6 4 ETHERNET Connector 64 6 5 P Connectors 65 6 6 UT PHASED ARRAY Connector 66 List of Figures 67 List of Tables 69 Index...

Страница 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...

Страница 7: ...y related labels and symbols are attached to the instrument at the locations shown in Figure i 1 on page 1 If any or all of the labels or symbols are missing or illegible please contact Olympus Figure...

Страница 8: ...pean Community See the Declaration of Conformity for details Contact your Olympus representative for more information The WEEE symbol indicates that the product must not be disposed of as unsorted mun...

Страница 9: ...nnectors shown in the figure below warns of this electric shock risk SERIAL The serial number is a nine 9 digit number in the following format yynnnddmm where yy Production year nnn Unit number manufa...

Страница 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...

Страница 11: ...ended use It must never be used to inspect or examine human or animal body parts Instruction Manual This instruction manual contains essential information on how to use this Olympus product safely and...

Страница 12: ...human injury and or equipment damage do not disassemble modify or attempt to repair the instrument Safety Symbols The following safety symbols might appear on the instrument and in the instruction ma...

Страница 13: ...performed or adhered to could result in death or serious personal injury Do not proceed beyond a WARNING signal word until the indicated conditions are fully understood and met The CAUTION signal word...

Страница 14: ...gory II Before turning on the instrument verify that the correct safety precautions have been taken see the following warnings In addition note the external markings on the instrument which are descri...

Страница 15: ...or short circuited fuse holders doing so could cause electric shock or create a fire hazard If there is any possibility that the ground protection could be impaired you must make the instrument inope...

Страница 16: ...accordance with the manufacturer s instructions may cause interference The TomoScan FOCUS LT has been tested and found to comply with the limits for an industrial device in accordance with the specif...

Страница 17: ...s intended solely as a teaching aid and shall not be used in any particular application without independent testing and or verification by the operator or the supervisor Such independent verification...

Страница 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...

Страница 19: ...lel Key Features TomoScan FOCUS LT instrument s key features include 16 32 or 64 focusing channels Up to 128 channels with linear amplifier for phased array inspection Real time data compression High...

Страница 20: ...plete description of the different TomoScan FOCUS LT model numbers please refer to Operating Specifications on page 42 Table 2 Possible TomoScan FOCUS LT configurations 16 focusing channels 32 focusin...

Страница 21: ...les the user to perform the following Connect a phased array probe to the TomoScan FOCUS LT Connect up to 4 conventional ultrasound probes Enable or disable UT conventional ultrasound mode Connect the...

Страница 22: ...N This connector allows the TomoScan FOCUS LT to be supplied from a DC power source The DC IN connector accepts voltages in the range of 15 VDC to 24 VDC rated at 5 A The external DC power supply must...

Страница 23: ...l ultrasound mode with the P125 P126 P127 and P128 connectors the voltage present on the P connectors can be dangerous and create a shock hazard AC The letters AC stand for alternating current When th...

Страница 24: ...econd it indicates that the unit is starting up When this indicator turns off after blinking it indicates that the unit is ready for operation ETHERNET This 8 position modular connector enables the co...

Страница 25: ...the AC source first disconnect the power cord from the AC outlet and then disconnect the power cord from the back of the instrument Failure to do so can cause electric shocks Handles Two handles are...

Страница 26: ...ure 1 3 on page 20 shows the back panel of the TomoScan FOCUS LT This panel enables the user to perform the following Connect the TomoScan FOCUS LT to an AC power source Replace the fuse Ground the un...

Страница 27: ...power source The DC IN connector located on the front panel accepts voltages between 15 VDC and 24 VDC rated at 5 A POWER CORD PROVIDES AC POWER DISCONNECT This marking on the back panel indicates tha...

Страница 28: ...nspection Fans Two fans cool the internal components Each fan has a mesh filter used to filter air drawn in by the fan Two grills for the air outlet are located on both sides of the TomoScan FOCUS LT...

Страница 29: ...bumpers of the back panel Do not leave the TomoScan FOCUS LT vertical on the bumpers of the back panel because the high center of gravity could make it unstable and likely to fall over Moreover the i...

Страница 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...

Страница 31: ...e packing list should include the following items TomoScan FOCUS LT with the proper internal configuration AC power cord Crossover Ethernet cable category 5e or higher USB drive including the TomoView...

Страница 32: ...e bumpers of the back panel because the high center of gravity could make it unstable and likely to fall over Moreover the insufficient clearance of the fans would impede proper cooling The TomoScan F...

Страница 33: ...c components or cause electric shocks To connect the TomoScan FOCUS LT instrument 1 Ensure that the TomoScan FOCUS LT is disconnected from the power source To prevent the risk of an electric shock fir...

Страница 34: ...ower outlet Turn on the TomoScan FOCUS LT by pressing the front panel on off switch 7 When using a DC power source connect the power cord to the DC IN connector of the TomoScan FOCUS LT front panel Co...

Страница 35: ...emotely turn on a TomoScan FOCUS LT from an AC power source When this mode is enabled you do not need to press the front panel on off switch to start the TomoScan FOCUS LT The instrument starts automa...

Страница 36: ...h on the front panel 3 Reconnect the DC power cord to the DC IN input connector or turn on the external DC power source 4 Release the on off switch on the front panel when the STBY indicator light is...

Страница 37: ...ation controls the acquisition process and analyzes the ultrasonic data collected by the instrument 3 1 Hardware Configuration Figure 3 1 on page 31 shows the hardware architecture of the TomoScan FOC...

Страница 38: ...in order to produce data UT I O The phased array section manages the focal laws transmission delay reception delay signal summing The RPC server is used for the configuration of the focal laws 3 3 Wo...

Страница 39: ...FOCUS LT is designed to be used with a 100BASE T Fast Ethernet network The TomoScan FOCUS LT has no internal disk drive The instrument must therefore be linked through an Ethernet network to a worksta...

Страница 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...

Страница 41: ...preventive maintenance Only a regular inspection of the instrument is recommended to ensure that the TomoScan FOCUS LT has adequate grounding The TomoScan FOCUS LT must always be completely inserted...

Страница 42: ...he connector contacts could stay wet and produce a short circuit when connecting cables 3 To remove persistent stains use a damp cloth with a soft soapy solution Do not use abrasive products or powerf...

Страница 43: ...lean the filter on both sides 4 Place the filter back in the appropriate position 5 Screw the four knurled screws back into place Figure 4 1 Fan filter removed from the TomoScan FOCUS LT 4 3 2 Changin...

Страница 44: ...qualified technician for replacement To change the fuse 1 Ensure that the instrument is turned off and the AC power cord is disconnected from the back panel of the instrument To prevent the risk of an...

Страница 45: ...ded could cause a fire hazard 5 Reinsert the fuse holder The fuse is now replaced and the TomoScan FOCUS LT should be ready for operation If the instrument still does not operate after the fuse replac...

Страница 46: ...ded and suggested by Olympus Table 4 Troubleshooting guide Problem Cause Solution The instrument does not turn on The fuse is blown Verify the fuse and replace it if applicable with a fuse of the corr...

Страница 47: ...ifications Category Specification Value Housing Size W H D 31 3 cm 11 4 cm 37 7 cm 12 32 in 4 49 in 14 83 in Net weight 5 kg 11 lb Environment Storage temperature 20 C to 60 C Operating temperature In...

Страница 48: ...up 5 2 Operating Specifications This section details the operating specifications of the TomoScan FOCUS LT 5 2 1 Pulser Receiver Differences and Limitations The pulser receiver specifications are give...

Страница 49: ...higher voltage in the load compared to models R Models with the letter R For law in pitch and catch mode Number of pulsing channels Maximum active channels Number of receiving channels Maximum active...

Страница 50: ...FLT16128 P O K 1 K L Pulse echo N 15 M 15 K 1 K L Pitch and catch N 7 M 7 Limit 4 K L M 16 n 1 2 3 4 5 6 7 FLT16128 R O K 1 K L Pulse echo N 15 M 15 K 1 K L Pitch and catch N 15 M 15 FLT32128 P O K 1...

Страница 51: ...1 FLT64128 R O K 1 K L Pulse echo N 63 M 63 K 1 K L Pitch and catch N 63 M 63 FLT6464 P O K 1 K L Pulse echo N 63 M 63 K 1 K L Pitch and catch N 31 M 31 Table 7 Specifications of models FLT16128 P O...

Страница 52: ...frequency at 3 dB Overall system bandwidth in pitch and catch 0 54 MHz 10 to 20 MHz 12 cuto frequency at 3 dB Filters All filters are 10 cuto frequency at 3 dB 0 4 MHz to 1 7 MHz band pass 1 1 MHz to...

Страница 53: ...s steps of 2 5 ns Dynamic depth focusing DDF Yes Table 8 Specifications of models FLT32128 P O and FLT32128 R O Category Specification FLT32128 P O FLT32128 R O Pulser Pulse voltage 40 V 80 V and 115...

Страница 54: ...andwidth 0 54 MHz 10 to 22 MHz 12 cuto frequency at 3 dB Filters All filters are 10 cuto frequency at 3 dB 0 4 MHz to 1 7 MHz band pass 1 1 MHz to 3 4 MHz band pass 2 0 MHz to 7 4 MHz band pass 3 3 MH...

Страница 55: ...s steps of 2 5 ns Dynamic depth focusing DDF Yes Table 9 Specifications of models FLT64128 P O and FLT64128 R O Category Specification FLT64128 P O FLT64128 R O Pulser Pulse voltage 40 V 80 V and 115...

Страница 56: ...ndwidth 0 54 MHz 10 to 23 5 MHz 12 cuto frequency at 3 dB Filters All filters are 10 cuto frequency at 3 dB 0 4 MHz to 1 7 MHz band pass 1 1 MHz to 3 4 MHz band pass 2 0 MHz to 7 4 MHz band pass 3 3 M...

Страница 57: ...0 10 s steps of 2 5 ns Dynamic depth focusing DDF Yes Table 10 Specifications of models FLT3232 P O and FLT6464 P O Category Specification FLT3232 P O FLT6464 P O Pulser Pulse voltage 50 V 100 V and 2...

Страница 58: ...3 4 MHz band pass 2 0 MHz to 7 4 MHz band pass 3 3 MHz to 7 5 MHz band pass 4 4 MHz to 11 MHz band pass 5 5 MHz to 15 MHz band pass 6 3 MHz to 20 MHz band pass 9 7 MHz to 22 MHz band pass 4 1 MHz to...

Страница 59: ...ond to the last four channels of the UT PHASED ARRAY connector The specifications for these channels are the same ones as the specifications for the phased array channels When the instrument is used i...

Страница 60: ...external signal Synchronization modes Pulser surface echo external trigger A scan recording Up to 8000 A scans s 512 samples of 8 bits C scan data Up to 20 kHz four measurement gates plus synchronizat...

Страница 61: ...quadrature clock directions up down Digital input 2 digital inputs for more information see I O Connector on page 60 Digital output 4 digital outputs for more information see I O Connector on page 60...

Страница 62: ...with the essential requirements of the applicable directives of the European Union that is the Low Voltage Directive 2006 95 EC and the EMC Directive 2004 108 EC The conformity has been assessed with...

Страница 63: ...e FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can rad...

Страница 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...

Страница 65: ...following information a brief description the manufacturer number the number of the corresponding cable connector an illustration and a table giving the specifications or the signal pinouts for the co...

Страница 66: ...ure 6 1 The I O connector Table 13 Pinout for the I O connector Pin I O Signal Description Type Signal length A Input PhB 1 Encoder 1 phase B direction TTLa B Output Dout1 Programmable output Configur...

Страница 67: ...ut Not programmed in TomoView TTLa P Input PhB 2 Encoder 2 phase B direction TTLa R Input Din2 Programmable input Can be configured as rotation sync input revolution synchronization signal or as prese...

Страница 68: ...required DIN signal is greater than 50 s then maximum PRF will be lower than 10 kHz Figure 6 2 The Trig on Dout 2 for a signal of 50R 6 2 DC OUT Connector Description 5 pin female connector Manufactur...

Страница 69: ...2 pin male connector Manufacturer number W W Fischer Inc D 103 Z051 60 Olympus 21AB0321 U8902550 Suggested cable connector number W W Fischer Inc S 103 Z051 130 Olympus 21AB0322 U8908014 Table 14 Pin...

Страница 70: ...nector Manufacturer number Stewart Connector SS 650810SAFLS Olympus 21AI0080 Suggested cable connector number Stewart Connector 940SP 360808 Olympus 21AI0079 Figure 6 5 The ETHERNET connector Table 15...

Страница 71: ...Suggested cable connector number LEMO FFC 00 250 CTAC31 Equivalent W W Fischer Inc S 101 A004 3 0 Olympus 21AB0016 U8907038 Figure 6 6 The P connectors Table 16 Pinout for the ETHERNET connector Pin...

Страница 72: ...UT PHASED ARRAY Connector Description 160 pin female Minidock type connector Manufacturer number Global Components 30033 160T Olympus 21AI0170 U8904017 Suggested cable connector number Framatome 89649...

Страница 73: ...e architecture of the TomoScan FOCUS LT 31 Figure 4 1 Fan filter removed from the TomoScan FOCUS LT 37 Figure 4 2 Pulling out the fuse holder from the TomoScan FOCUS LT 39 Figure 4 3 Fuse and spare fu...

Страница 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...

Страница 75: ...s of models FLT16128 P O and FLT16128 R O 45 Table 8 Specifications of models FLT32128 P O and FLT32128 R O 47 Table 9 Specifications of models FLT64128 P O and FLT64128 R O 49 Table 10 Specifications...

Страница 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...

Страница 77: ...8 P O 51 FLT64128 R O 51 FLT6464 P O 52 Bootp server 33 bumpers protective 19 22 button UT ON 16 C cable part number crossover 33 Canadian and U S standards 56 CAUTION signal word 7 cautions See also...

Страница 78: ...EMC 56 disposal equipment 10 E electric shock danger note 3 EMC directive 56 EMC directive compliance 10 environment specifications 41 equipment disposal 10 Ethernet cable 55 56 connector 18 64 65 li...

Страница 79: ...36 compatibility 5 connecting 27 description 13 overview 15 back panel 20 front panel 15 positioning warning 23 26 K key features 13 L labels 1 LEDs See indicator lights LEMO 01 connectors 3 link Ethe...

Страница 80: ...supply warning 19 21 precautions battery 9 preventive maintenance 35 probes connecting 28 caution 28 protective bumpers 19 22 pulser receiver specifications 42 FLT16128 P O 45 FLT16128 R O 45 FLT32128...

Страница 81: ...uipment 25 system specifications 41 T TCG specifications 54 technical support 11 terminal ground 22 Time Corrected Gain specifications 54 TIP signal word 8 troubleshooting 40 changing the fuse 38 U U...

Страница 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...

Отзывы: