![Olympus TomoScan FOCUS LT Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/olympus/tomoscan-focus-lt/tomoscan-focus-lt_user-manual_3891023046.webp)
DMTA041-01EN [U8778479], Rev. C, May 2014
Chapter 4
40
Figure 4-3 Fuse and spare fuse locations (fuse holder pulled out - side view)
4.5
Troubleshooting
This section will help you solve minor problems that could occur during the operation
of your TomoScan FOCUS LT. Olympus has developed this troubleshooting guide
assuming that the instrument has not been modified, and that the cables and
connectors are those provided and suggested by Olympus.
Table 4 Troubleshooting guide
Problem
Cause
Solution
The instrument does
not turn on.
The fuse is blown.
Verify the fuse and replace it, if
applicable, with a fuse of the
correct rating. For the detailed
procedure, see “Fuse Changing”
on page 38.
Communication
problems occur
between the
computer and the
TomoScan
FOCUS LT.
The workstation and
the TomoScan
FOCUS LT
are not
connected with the
proper Ethernet cable
type.
Ensure that you use a crossover
Ethernet cable for a direct
connection between the
workstation and the TomoScan
FOCUS LT.
Location of spare fuse
Location of fuse in use
Содержание TomoScan FOCUS LT
Страница 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...
Страница 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...
Страница 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...
Страница 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...
Страница 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...
Страница 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...
Страница 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...
Страница 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...
Страница 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...