11
SZ2 series
Ensuring the confocality between the observation image and
the camera monitor image makes it possible to maintain
accurate focusing of the observation image or monitor image
even when the light path is switched.
Adjust the diopter of the eyepiece (see page 7) and bring the
specimen into focus.
Pull out the light path selector lever a and set the magnification to
the minimum.
Slightly loosen the two confocality adjustment clamping screws b
and C-mount clamping screw c .
Hold the camera d and rotate the confocality e to focus on the
monitor image.
After the adjustment, tighten the clamping screws b and c .
Applicable cameras (image pickup device size) and field numbers
The camera light path incorporates a 0.5X lens.
TIP
1
2
3
4
TIP
a
b
c
d
e
2/3 in. image pickup device
(Diagonal length 22 mm)
1/2 in. image pickup device
(Diagonal length 16 mm)
FN 22
(With 10X eyepiece)
Rotation of monitor image
The monitor image is rotated slightly because the right observation
light path, which is tilted slightly, is used as the camera light path.
The monitor image can be aligned with the observation image by
rotating the camera.
8
Adjusting the Confocality of the Camera (SZ61TR)
Downloaded from
Содержание SZ61
Страница 2: ...Downloaded from ManualsNet com search engine...
Страница 4: ...Downloaded from ManualsNet com search engine...