background image

– 5 –

12

Place the air jack between the casters against the caster on the rear side. At this time, confirm that the band of air jack 
is not removed.

13

Close the air outlet plug (pink) of the air jack by rotating the plug 
clockwise.

Figure 12

14

Take out the pump from the package carton, remove the cap of the air inlet plug (white) and supply air until stand R can 
be removed.

CAUTION

Supply air in with the air jack in contact with the caster. Otherwise, the product may topple down.

Band

Figure 11

Front

Rear

Air outlet plug (pink)

Pump

Air inlet plug 
(white)

Air jack

Pump

Figure 13

Содержание ORBEYE OME-V200

Страница 1: ...se the product may topple down or fall from the pallet NOTE Recommend that you work with two or more people For unpacking a phillips screwdriver cutter nipper is required The side that is described as...

Страница 2: ...e nipper CAUTION Be careful not to cut your hand when cutting the ester band 2 Remove the cap Figure 3 NOTE The removed cap will later be used as a slope 3 Cut the tape of four corners of the pallet b...

Страница 3: ...phillips screwdriver Figure 5 5 While holding the sleeves so that the sleeves will not collapse remove the sleeves 2 6 Pull and remove the pads 8 7 Remove the BI pad Figure 7 Metallic fasteners Sleeve...

Страница 4: ...arton Figure 8 9 While holding the frame so that the frame will not collapse remove the pipes 6 Figure 9 10 Remove the frame 11 Loosen the strap of bag 2 remove bag 2 Package carton Pipes Frame Bag 1...

Страница 5: ...air jack by rotating the plug clockwise Figure 12 14 Take out the pump from the package carton remove the cap of the air inlet plug white and supply air until stand R can be removed CAUTION Supply air...

Страница 6: ...et plug pink of the air jack by rotating the plug counterclockwise until the air jack can be moved Figure 16 CAUTION When exhausting air be careful not to have your hand or an object caught by the cas...

Страница 7: ...of the air inlet plug white and supply air until stand F can be removed using the pump CAUTION Supply air in with the air jack in contact with the caster Otherwise the product may topple down NOTE Rem...

Страница 8: ...ir be careful not to have your hand or an object caught by the caster Exhaust air slowly by gradually opening the plug If it is opened abruptly the product may fall suddenly 22 Remove the air jack 23...

Страница 9: ...Figure 26 CAUTION Be careful not to have your foot caught under this instrument or caster Otherwise foot injury may result Do not put your hand below this instrument Otherwise hand injury may result...

Страница 10: ...ing that this instrument is locked to the floor and then release your hand from movement handles Otherwise this instrument will have momentum and fall over 29 Remove the cable protection cap on the re...

Страница 11: ...11 NOTE During and after removal of bag 1 be careful not to touch the lens surface on the microscope section Figure 31 Microscope section Lens surface...

Страница 12: ...3 Moscow Russia Fax 7 495 663 8486 Telephone 7 495 926 7077 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A Fax 484 896 7128 Telephone 484 896 5000 5301 Blue Lagoon Drive Suite 29...