Chapter 4 Operation
BF TYPE 160 Series OPERATION MANUAL
Ultrasonic observation
• When withdrawing the ultrasonic probe with balloon sheath
from the endoscope, make sure that the balloon is
completely deflated. Withdrawing the probe while the balloon
is inflated could result in patient injury and/or could damage
the ultrasonic probe.
• When using the ultrasonic probe with balloon sheath, always
lubricate the balloon before inserting the balloon sheath into
the instrument channel. Otherwise, the balloon could rupture
or come off. This could result in patient injury.
Prepare, inspect and connect the ultrasonic observation unit and ultrasonic
probe as described in their instruction manuals.
The ultrasonic probe with balloon sheath can be used in
combination with the BF-1T160/XT160.
Bronchoalveolar lavage
Using the BAL (bronchoalveolar lavage) kit
1.
Disconnect the suction tube from the suction valve. Connect the suction
tube to the suction connector of a commercially available BAL kit.
Connect the BAL kit’s suction line to the suction connector of the suction
valve (see Figure 4.5).
Figure 4.5
2.
Securely insert a syringe filled with lavage fluid (e.g. saline) into the slit of
the biopsy valve and press the plunger to feed lavage fluid.
3.
Press the suction valve to aspirate lavage fluid.