28
Chapter 3 Preparation and Inspection
BF TYPE 160 Series OPERATION MANUAL
1.
Inspect the sterile package for tears, inadequate sealing or water damage. If
the sterile package shows any irregularities, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2.
Confirm that the button can be pushed without excessive resistance.
3.
Inspect the suction valve for cracks or other damage (see Figure 3.3).
Figure 3.3
Inspection of the single-use biopsy valve (MAJ-210)
• The biopsy valve is provided in a sterile condition. Do not
open the package until you are ready to use it.
• Do not use an instrument after the expiration displayed on
the sterile package. Doing so may pose an infection control
risk.
• Do not attempt to reuse or sterilize the biopsy valve. This
could pose an infection control risk or cause equipment
damage.
This valve is designed for use with BF-30, 40, 200, 240, and
160 series endoscopes.
Suction
connector
Button