FLUSHING PUMP OFP

3

2

O

PERATING

P

RECAUTIONS

2.1

Explosion hazard -- never install or use the OFP within the zone of risk of flammable gases.

2.2

The OFP should only be used in a medical facility under the direction of a trained physician.

2.3

The operator must assess the condition of the patient and use clinical judgement to set the flow rate from the pump to a
suitable level to avoid patient trauma. The flow control should always be set to MIN at the start of the procedure and be
increased progressively to a level commensurate with the clinical condition of the patient and degree of washing required.

Note that when the flow control is set to MAX, the flow rate through the washing channel of the endoscope will be in
excess of that which can be delivered when using a 50ml syringe.

2.4

Use only in conjunction with Olympus EVIS EXERA series endoscopes having an auxiliary water inlet.

2.5

Only use the Olympus auxiliary water tube (MAJ--855) and irrigation tube supplied to connect the OFP to the endoscope,
otherwise patient safety may be compromised or damage to the pump may occur.

2.6

Note that the luer lock connector on the MAJ--855 which attaches to the OFP filter includes a one--way valve to prevent
backflow -- do NOT use the MAJ--855 without this connector in place, or irrigation tube and filter, container contamination
and damage may result.

2.7

The irrigation tube and filter MUST be autoclaved before use.

2.8

The irrigation tube and filter MUST be replaced daily -- failure to do so may lead to patient infection and a reduction in
equipment performance.

2.9

Reprocess the MAJ--855 following each use according to the manufacturer’s instructions supplied with the product.

2.10 Check the “USE BY” date of the tubing as indicated on the packaging label. Do not use any tubes beyond the specified

“USE BY” date.

2.11 Never allow the fluid container to run dry, otherwise air will be supplied to the patient.
2.12 To prevent free--flow of fluid to the patient, do not open the pump cover when the OFP is connected to an endoscope which

is inserted into the patient and ensure the OFP is positioned below patient level.

2.13 If the filter becomes blocked or the irrigation tube from the pump to the fluid container is blocked or kinked, the flow rate will

be noticeably reduced. Check the tube for kinks or blockages and the filter for contamination, replace if necessary.

2.14 Should the OFP be operated with a blocked endoscope or filter, damage to the irrigation tube may occur. In this event, care

must be taken when disconnecting the fluid path as the system may be pressurised. To depressurise the system, the pump
head should be opened to allow fluid back into the bottle. This must only be carried out with the endoscope removed from
the patient. The filter and irrigation tube MUST be replaced before the unit is reused.

2.15 Always switch off the pump before fitting or replacing the irrigation tube.
2.16 When the irrigation tube has been correctly installed and the pump operated,do notdisturb the position ofthe irrigation tube

eg: by opening the pump head, otherwise the irrigation tube must be replaced. Failure to do so may lead to a reduction in
equipment performance.

2.17 Periodically inspect the MAJ--855 for obvious signs of wear or damage and replace if necessary or as recommended in the

manufacturers instructions.

2.18 Periodically inspect the fluid container for obvious signs of wear or damage and replace if necessary.
2.19 Always empty the fluid container before moving the equipment. Care should be taken to obtain a firm grip underneath the

body of the equipment before moving. The fluid container holder must never be used as a hand hold or carrying handle.

2.20 Store and use the OFP within the environmental conditions described in Section 8; failure to do so may lead to equipment

malfunction or failure.

2.21 Only use replacement parts as specified in Section 7. Failure to do so may lead to equipment malfunction or failure.

Содержание OFP

Страница 1: ...INSTRUCTIONS ENDOSCOPIC FLUSHING PUMP OFP CAUTION Federal USA law restricts this device to sale by or on the order of a physician 0086...

Страница 2: ...taltic flushing pump intended for use in medical facilities under the direction of a trained physician The OFP has been designed to be used with Olympus EVIS EXERA gastrointestinal videoscopes having...

Страница 3: ...G PRECAUTIONS 3 3 INSTRUCTIONS FOR USE 4 4 STEAM STERILIZATION AUTOCLAVING 9 5 CARE AND STORAGE 10 6 MAINTENANCE AND REPAIR 10 7 SPARES ACCESSORIES 12 8 SPECIFICATIONS 12 9 TECHNICAL DESCRIPTION 13 10...

Страница 4: ...he standard set are present Contact your Olympus service centre or nearest Olympus office if any parts are damaged or missing FIGURE 1 1 3 1 4 2 5 Contents Qty 1 Olympus Flushing Pump OFP 1 2 Fluid co...

Страница 5: ...ote control cable connection Label 3 IEC power receptacle with integral fuses Potential equalisation terminal Fluid container strap Fluid container holder 1 3 Labels Label 1 Refer to Instructions Labe...

Страница 6: ...e air will be supplied to the patient 2 12 To prevent free flow of fluid to the patient do not open the pump cover when the OFP is connected to an endoscope which is inserted into the patient and ensu...

Страница 7: ...tion of the patient and use clinical judgement to set the flow rate from the pump to a suitable level to avoid patient trauma The flow control should always be set to MIN at the start of the procedure...

Страница 8: ...rformance and will recover if left to rest 3 2 Place the OFP on a level flat surface below patient level ensuring the fluid container is not exposed to heat emitted from adjacent equipment such as a l...

Страница 9: ...d auxiliary water inlet on the endoscope CAUTION Use only in conjunction with Olympus EVIS EXERA series endoscopes having an auxiliary water inlet and MAJ 855 auxiliary water tube Ensure the MAJ 855 h...

Страница 10: ...delivered when using a 50ml syringe CAUTION NEVER allow the fluid container to run dry otherwise air will be supplied to the patient 3 12 To refill the fluid container during use leave the pump cover...

Страница 11: ...isconnect the endoscope luer connector Figure 3 5b iv allow any fluid in the tube to return to the fluid container then remove the tube from the container v discard the irrigation tubing and filter at...

Страница 12: ...ry Handling a wet package can compromise its sterility For details on operation of the autoclave refer to the instruction manual for the autoclave Typical parameters for pre vacuum autoclaves for wrap...

Страница 13: ...ead Empty the fluid container and autoclave prior to storage 2 The OFP should be stored within the environmental conditions given in Section 8 6 MAINTENANCE AND REPAIR 6 1 Routine Maintenance The foll...

Страница 14: ...eck the condition of both fuses located at the rear of the unit and the fuse in the power cable plug as appropriate CAUTION The fuse must always be replaced with one of the correct rating If the fuse...

Страница 15: ...ty This product complies with the requirements of EN IEC 60601 1 2 for emissions and immunity and as such its operation is unlikely to be affected by or cause interference with equipment meeting appro...

Страница 16: ...staltic Pump operation Water feed controlled by pneumatic footswitch Flow rate variable controlled from front panel Flow rate delivery from the pump will depend on the length and type of Olympus EVIS...

Страница 17: ...d Check tubing and replace if necessary with specified tube see Section 7 Pump head or tube rollers damaged Switch unit OFF open pump cover and check for signs of damage If damage is evident refer to...

Страница 18: ......

Страница 19: ......

Страница 20: ...01702 465677 Telephone 01702 616333 2951 Ishikawa cho Hachioji shi Tokyo 192 8507 Japan Fax 0426 46 2429 Telephone 0426 42 2111 OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS SINGAPORE PTE LTD OLYMPUS MOSCOW LI...

Отзывы: