background image

FLUSHING PUMP OFP

12

7

S

PARES & ACCESSORIES

NOTE

To maintain standards compliance, use only the following Spares and Accessories with this product.

7.1

Spares

Instructions

5070135

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Fluid container

7501352

. . . . . . . . . . . . . . . . . . . . . . . . . . . .

Fluid container strap

7501353

. . . . . . . . . . . . . . . . . . . . . . . .

Irrigation tube (pack of 10)

7501669

. . . . . . . . . . . . . . . . . . . .

Filter (pack of 10)

7501355

. . . . . . . . . . . . . . . . . . . . . . . . . .

Footswitch

7501357

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Fuse T1A (pack of 5)

7501356

. . . . . . . . . . . . . . . . . . . . . . .

Fuse T2A (pack of 5)

7501404

. . . . . . . . . . . . . . . . . . . . . . .

Power cable (UK)

7145454

. . . . . . . . . . . . . . . . . . . . . . . . . .

Power cable (USA)

7318766

. . . . . . . . . . . . . . . . . . . . . . . . .

Power cable (Europe)

7145462

. . . . . . . . . . . . . . . . . . . . . . .

Power cable (RoW)

7318561

. . . . . . . . . . . . . . . . . . . . . . . . .

MAJ--855 endoscope tube

1113054 (UK part number only)

. . . . . . . . . . . . . . . . . . . .

7.2

Accessories

Remote control cable MAJ--920

8

S

PECIFICATIONS

Items

Specifications

Product name

Olympus Flushing Pump OFP

Standards

compliance

This symbol on the 110--120V version indicates it is listed by

Intertek Testing Services as meeting the requirements of UL

2601--1 and CAN/CSA C22.2 601.1--M90.

3024005

Classification

Electro-magnetic

compatibility

This product complies with the requirements of EN IEC 60601-1-2 for emissions and

immunity, and as such, its operation is unlikely to be affected by, or cause interference

with, equipment meeting appropriate EMC standards. As a precaution, equipment which

may be sensitive to interference outside the limits specified by EN IEC 60601--1--2 should

not be placed in close proximity to the OFP.

Classification

(electromedical

equipment)

Type of protection

against electrical

shock

220-240V model

Classification according to EN IEC 60601-1: Class I Type BF.

110-120V model

Classification according to UL 2601-1: Class I Type BF

Degree of protection

against electrical

shock

In accordance with EN IEC 60601-1 and UL 2601--1, the OFP is marked with the

symbol

to indicate the provision of an adequate degree of protection against electrical

shock and that it has an applied part isolated from all other parts of the equipment.

Degree of protection

against explosion

None: the Olympus OFP must NOT be used in the zone of risk of flammable anaesthetic

gases.

Mode of operation

Continuous

Содержание OFP

Страница 1: ...INSTRUCTIONS ENDOSCOPIC FLUSHING PUMP OFP CAUTION Federal USA law restricts this device to sale by or on the order of a physician 0086...

Страница 2: ...taltic flushing pump intended for use in medical facilities under the direction of a trained physician The OFP has been designed to be used with Olympus EVIS EXERA gastrointestinal videoscopes having...

Страница 3: ...G PRECAUTIONS 3 3 INSTRUCTIONS FOR USE 4 4 STEAM STERILIZATION AUTOCLAVING 9 5 CARE AND STORAGE 10 6 MAINTENANCE AND REPAIR 10 7 SPARES ACCESSORIES 12 8 SPECIFICATIONS 12 9 TECHNICAL DESCRIPTION 13 10...

Страница 4: ...he standard set are present Contact your Olympus service centre or nearest Olympus office if any parts are damaged or missing FIGURE 1 1 3 1 4 2 5 Contents Qty 1 Olympus Flushing Pump OFP 1 2 Fluid co...

Страница 5: ...ote control cable connection Label 3 IEC power receptacle with integral fuses Potential equalisation terminal Fluid container strap Fluid container holder 1 3 Labels Label 1 Refer to Instructions Labe...

Страница 6: ...e air will be supplied to the patient 2 12 To prevent free flow of fluid to the patient do not open the pump cover when the OFP is connected to an endoscope which is inserted into the patient and ensu...

Страница 7: ...tion of the patient and use clinical judgement to set the flow rate from the pump to a suitable level to avoid patient trauma The flow control should always be set to MIN at the start of the procedure...

Страница 8: ...rformance and will recover if left to rest 3 2 Place the OFP on a level flat surface below patient level ensuring the fluid container is not exposed to heat emitted from adjacent equipment such as a l...

Страница 9: ...d auxiliary water inlet on the endoscope CAUTION Use only in conjunction with Olympus EVIS EXERA series endoscopes having an auxiliary water inlet and MAJ 855 auxiliary water tube Ensure the MAJ 855 h...

Страница 10: ...delivered when using a 50ml syringe CAUTION NEVER allow the fluid container to run dry otherwise air will be supplied to the patient 3 12 To refill the fluid container during use leave the pump cover...

Страница 11: ...isconnect the endoscope luer connector Figure 3 5b iv allow any fluid in the tube to return to the fluid container then remove the tube from the container v discard the irrigation tubing and filter at...

Страница 12: ...ry Handling a wet package can compromise its sterility For details on operation of the autoclave refer to the instruction manual for the autoclave Typical parameters for pre vacuum autoclaves for wrap...

Страница 13: ...ead Empty the fluid container and autoclave prior to storage 2 The OFP should be stored within the environmental conditions given in Section 8 6 MAINTENANCE AND REPAIR 6 1 Routine Maintenance The foll...

Страница 14: ...eck the condition of both fuses located at the rear of the unit and the fuse in the power cable plug as appropriate CAUTION The fuse must always be replaced with one of the correct rating If the fuse...

Страница 15: ...ty This product complies with the requirements of EN IEC 60601 1 2 for emissions and immunity and as such its operation is unlikely to be affected by or cause interference with equipment meeting appro...

Страница 16: ...staltic Pump operation Water feed controlled by pneumatic footswitch Flow rate variable controlled from front panel Flow rate delivery from the pump will depend on the length and type of Olympus EVIS...

Страница 17: ...d Check tubing and replace if necessary with specified tube see Section 7 Pump head or tube rollers damaged Switch unit OFF open pump cover and check for signs of damage If damage is evident refer to...

Страница 18: ......

Страница 19: ......

Страница 20: ...01702 465677 Telephone 01702 616333 2951 Ishikawa cho Hachioji shi Tokyo 192 8507 Japan Fax 0426 46 2429 Telephone 0426 42 2111 OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS SINGAPORE PTE LTD OLYMPUS MOSCOW LI...

Отзывы: