34
Reprocessing
- Telescope:
- Neodisher MediClean forte manufactured by Dr. Weigert
- Neodisher Z manufactured by Dr. Weigert
- Protective tube:
- Sekumatic FR manufactured by Henkel-Ecolab
- Neodisher Z manufactured by Dr. Weigert
7.6 Maintenance
Inspecting the telescope
• Visually inspect the product thoroughly after cleaning. The
product must be visually clean. If there are any signs of debris,
repeat the cleaning process.
• Observe the instructions in the section “Inspection” on page
19.
Inspecting the protective tube
• Visually inspect the product thoroughly after cleaning. The
product must be visually clean. If there are any signs of debris,
repeat the cleaning process.
• Check that the product has no dents, cracks, bending, or
deformations.
• Check all markings on the product for clear visibility.
Even products designed to be reused have a limited service
life. A number of factors connected with handling and some
reprocessing methods may lead to increased wear of the product.
The service life can be markedly shortened. The product must be
replaced if signs of wear become visible.
7.7 Sterilization
Sterilization of the product has been validated with steam
sterilization and STERRAD
®
sterilization.
• Choose one of the sterilization methods.
• For steam sterilization, refer to the section ”Steam sterilization”
on page 35.