24
Use
• Do not use high intensity illumination in cavities with small
lumens.
• Do not use high intensity illumination when the distal end of the
endoscope is placed into close proximity to the tissue.
• If the distal end is soiled and impairs proper visualization,
remove the endoscope and clean the distal end.
CAUTION
Risk of injury to the user
The light-guide connector on the endoscope becomes hot. There
is a risk of burns if the endoscope is held by hand for a longer
period.
• Avoid touching the light-guide connector when holding the
endoscope.
NOTICE
Risk of damage to the product
If activated laser equipment touches the energized endoscope,
the energized endoscope may be damaged.
• Only ignite when the laser emission part is visible on the
endoscopic image.
5.2 Procedure
Using PDD mode
Olympus recommends to use PDD mode for diagnosis only and
white light imaging during treatment.