36
Glare Prevention Setting
(See page 20.)
1FPV 1FPV?
Functions
1. Sets the glare prevention.
2. Cancels the glare prevention.
Special notes
1. The glare prevention is set or canceled according to the setup of DIP switch No. 4 at the moment the microscope is
turned ON. This command is used to change this initial condition.
2. The glare prevention setting of the MX61/61L made using instruction command 1FPV ON is given priority over the DIP
switch setting. The setting using 1FPV ON is enabled even after the microscope is switched the remote mode to local
mode, until the microscope is turned OFF.
Command descriptions
1FPV ON
Request
Setting the anti-glare setting.
(in darkfield observation only)
1FPV +
Setting is complete.
1FPV !,E01120
A string other than ON/OFF is
sent after “1FPV ”.
1FPV OFF
Command
Type
Description
Response
Description
Request
Cancellation of the anti-glare
setting.
1FPV +
Cancellation is complete.
1FPV !,E01120
A string other than ON/OFF is
sent after “1FPV ”.
1FPV?
Query
Check of the current glare
prevention setting status.
1FPV ON
Glare prevention is ON.
1FPV OFF
Glare prevention is OFF.
1FPV !,E01150
A string is sent after “1FPV? ”.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.