16
SZX10, SZX7
Loosen the clamping knob B of LG-R66 (Ring light guide) A .
Insert SZX-LGR66 (Ring light guide adapter for SZX) C into
LG-R66 A completely to the end and tighten the clamping knob B .
Loosen the clamping screw D of SZX-LGR66 C sufficiently using
the Allen screwdriver provided with the microscope frame.
Insert SZX-LGR66 C into the objective of SZX, align the surface of
objective with the bottom surface of LG-R66, and tighten the clamping
screw D using the Allen screwdriver.
1
2
3
4
3
Attaching LG-R66 (Ring light guide)
1
3
B
A
C
SZ61, SZ51
Screw in SZ-LGR66 (Ring light guide adapter for SZ) A completely to
the thread of the auxiliary objective of the SZ2 microscope.
Loosen the clamping knob C of LG-R66 (Ring light guide) B .
Insert LG-R66 B into SZ-LGR66 A completely to the end and tighten
the clamping screw C .
1
2
3
SZX16
Loosen the clamping knob B of LG-R66 (Ring light guide) A .
Insert LG-R66 A upward until it contacts the objective bottom surface
C , and tighten the clamping knob B .
1
2
A
A
B
B
C
C
D
2
2
4