Chapter 5 Troubleshooting
67
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
EndoTherapy accessories
Other
Irregularity
description
Possible cause
Solution
EndoTherapy
accessory does not
pass through the
instrument channel
smoothly.
An incompatible
EndoTherapy accessory is
being used.
Refer to the “System chart” in the
Appendix and select a compatible
EndoTherapy accessory. Confirm
that the color code on the
EndoTherapy accessory matches
that on the endoscope.
Guidewire cannot
be locked.
Guidewire is not locked at
its stiff part.
Lock the guidewire at its stiff part.
Guidewire with a diameter
less than ø 0.64 mm is
used.
Select a guidewire with a diameter
of ø 0.64 mm or more.
Guidewire-locking groove is
dirty.
Clean and disinfect or sterilize the
guidewire-locking groove as
described in the “REPROCESSING
MANUAL” with your endoscope
model listed on the cover.
Contrast media is
congealed in the guidewire
lumen of the EndoTherapy
accessory.
Clean the lumen of the
EndoTherapy accessory and then
insert/withdraw it.
Irregularity
description
Possible cause
Solution
The remote switch
does not work.
The wrong remote switch is
operated.
Operate the correct remote switch.
The remote switch function
has been set incorrectly.
Set the remote switch function
correctly as described in the video
system center’s instruction manual.