
56
Chapter 4 Operation
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
• When the guidewire is placed into the biliary or pancreatic
duct with papilla observed in the left or right side of the
endoscopic image, the guidewire may move outside the view
of the endoscopic image because the forceps elevator is
raised extensively. In this case, do not operate the bending
section or insert or withdraw the insertion section forcibly or
abruptly. Patient injury, bleeding, and/or perforation may
result. If the guidewire moves outside the view of the
endoscopic image, perform treatment carefully while
observing the X-ray image, or lower the forceps elevator and
locate the papilla as centrally as possible in the endoscopic
image by adjusting the position of the distal end of
endoscope, and then raise the forceps elevator again.
• The assist function of the guidewire locking works most
effectively with guidewires with a diameter of ø 0.64 mm
(0.025 inch) or more.
• The assist function of the guidewire locking may not work
effectively due to various shapes and sizes of the patient’s
duodenum, biliary duct, and pancreatic duct.
• The assist function of the guidewire locking may not work
effectively under the following conditions:
If the elevator control lever is not held stationary.
If the proximal ends of the wire-guided type EndoTherapy
accessory and the guidewire are not straight.
If the contrast media in the guidewire lumen of the
EndoTherapy accessory is not washed with saline
solution.
If the wire-guided type EndoTherapy accessory is kinked,
deformed or damaged.
If the combination of the guidewire and the wire-guided
type EndoTherapy accessory is incorrect.
If the guidewire is not inserted sufficiently into the
biliary/pancreatic duct.
If an attempt is made to lock more than one guidewire
simultaneously.
If the position of the distal end of the endoscope and the
papilla is not appropriate for the assist function of the
guidewire locking (refer to “NOTE” on page 60).
Содержание EVIS EXERA II
Страница 2: ......