Chapter 3 Preparation and Inspection
35
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
Connection of the endoscope and ancillary equipment
Firmly connect the suction tube from the suction pump to the
suction connector on the endoscope connector. If the suction
tube is not attached properly, debris may drip from the tube
and can pose an infection control risk, cause equipment
damage, and/or reduce suction capability.
1.
If any ancillary equipment is ON, turn it OFF.
2.
Insert the endoscope connector completely into the output socket of the light
source.
3.
Place the water container’s water supply channel onto the water supply
connector on the endoscope connector at an angle of 90
and push it in until
it stops (see Figure 3.14 (1)).
4.
Turn the water container’s connection adapter 90
clockwise to align the air
supply channel with the air supply connector on the endoscope connector
(see Figure 3.14 (2)).
5.
Push the water container’s connection adapter again until it stops (see
Figure 3.14 (3)).
6.
Confirm that the water container’s connection adapter fits properly and that
it cannot be rotated (see Figure 3.14 (4)).
Figure 3.14
(1)
(2)
(3)
(4)
Water supply connector
Air supply connector
Water container’s connection adapter
Endoscope connector
Air supply channel
Water supply channel
Metal tip