Chapter 5 Troubleshooting
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Flexibility adjustment
(for endoscopes with flexibility adjustment only)
Auxiliary water feeding
(for endoscopes with auxiliary water feeding only)
Endo-therapy accessories
Others
Irregularity
description
Possible cause
Solution
Too difficult to
turn the
flexibility
adjustment ring.
The insertion tube is
looped.
Straighten the insertion tube.
Irregularity
description
Possible cause
Solution
The auxiliary
water inlet cap
is leaking.
The auxiliary water
inlet cap is worn out.
Replace it with a new one.
The auxiliary water
inlet cap is incorrectly
installed.
Install the auxiliary water inlet cap correctly.
Irregularity
description
Possible cause
Solution
Endo-therapy
accessory does
not pass
through the
instrument
channel
smoothly.
An incompatible
endo-therapy
accessory is being
used.
Refer to the “System chart” in the Appendix
and select a compatible endo-therapy
accessory. Confirm that the color code on
the endo-therapy accessory matches that
on the endoscope.
Irregularity
description
Possible cause
Solution
The remote
switch does not
work.
The wrong remote
switch is operated.
Operate the correct remote switch.
The remote switch
function has been set
improperly.
Set the remote switch function correctly as
described in the video system center’s
instruction manual.