Chapter 4 Operation
63
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
• To prevent the patient from accidentally biting the insertion
tube during an examination, it is strongly recommended that
a mouthpiece be placed in the patient’s mouth before
inserting the endoscope (for GIF models only).
• To prevent the patient from accidentally loosing dental
prosthesis, make sure that the patient removes them before
the examination.
• Do not apply olive oil or products containing petroleum-based
lubricants (e.g. vaseline). These products may cause
stretching and deterioration of the bending section’s
covering.
• Do not allow the insertion tube to be bent within a distance of
10 cm or less from the junction of the boot. Insertion tube
damage can occur (see Figures 4.2 and 4.3).
Figure 4.2
Figure 4.3
> 10 cm
Endoscopes with flexibility adjustment
> 10 cm