Chapter 5 Troubleshooting
81
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Suction
Image quality or brightness
Irregularity
description
Possible cause
Solution
The suction is
absent or
insufficient.
The biopsy valve is not
attached properly.
Attach it correctly.
The biopsy valve is
damaged.
Replace it with a new one.
The suction pump is
not set properly.
Adjust the suction pump’s setting as
described in its instruction manual.
The suction valve is
damaged.
Replace it with a new one.
The suction
valve is sticky.
The suction valve is
dirty.
Remove the suction valve. Reprocess the
suction valve and attach it again.
The suction valve is
damaged.
Replace it with a new one.
The suction
valve cannot be
attached.
The suction valve is
damaged.
Replace it with a new one.
An incorrect suction
valve is used.
Use a correct suction valve.
Liquid leaks out
from the biopsy
valve.
The biopsy valve is
damaged.
Replace it with a new one.
The biopsy valve is not
attached properly.
Attach it correctly.
Irregularity
description
Possible cause
Solution
There is no
video image.
Not all power switches
are ON.
Turn all the power switches ON.
An image is not
clear.
The objective lens is
dirty.
Feed water to remove mucus, etc.
An image is
excessively
dark or bright.
The light source is not
set properly.
Adjust the light source’s setting as
described in its instruction manual.
An image is
abnormal.
An incompatible video
system center is being
used.
Select a compatible video system center.
An incompatible light
source is being used.
Select a compatible light source.