background image

4

CX22LED

STANDARD COMPONENT UNITS

}After opening the package, make sure that the correct component units for the selected set are included.

The 100X objective and immersion oil are provided only with the CX22LEDRFS1 set.

The objectives have been screwed in tightly to prevent them from being loosened during transportation.
To remove an objective, turn it counterclockwise while holding it with a rubber sheet, etc., so that your
fingers don’t slip.

Eyepieces
(Fixed at 10X)

Stage

Condenser

Microscope frame

 · 4X, 10X, 40X, 100X
   (CX22LEDRFS1)
 · 4X, 10X, 40X
   (CX22LEDRFS2)

Immersion oil

 · Provided with a set including

the 100X objective.

     Optional Accessories

 · Filter Holder
   CH2-FH

 · Reflection Mirror
   CH20-MM

 · Eyepieces
   WHC15X
   WHC15X-H

 · Darkfield Ring
   CH2-DS

AC adapter

Power cord

Binocular observation
tube

Objectives

Revolving nosepiece

 · Cord Hanger
   CX22LED-CH

CAUTION

Содержание CX22LED

Страница 1: ...opes Model CX22LED To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before operating...

Страница 2: ...ations Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense FCC WARNING Chan...

Страница 3: ...nce 9 6 Adjusting the Condenser Position and Aperture Iris Diaphragm 10 7 Switching the Objectives 10 8 Using the 100X Immersion Objective 11 1 STANDARD COMPONENT UNITS 2 NOMENCLATURE 3 SUMMARY OF BRI...

Страница 4: ...he emission and immunity requirements described in IEC61326 series 7 The electromagnetic environment should be evaluated prior to operation of this product Do not use this product in close proximity t...

Страница 5: ...space around the AC adapter when installing the microscope 5 When carrying the microscope first make sure that the AC adapter has been disconnected and hold both sides around the hole of the arm as s...

Страница 6: ...r to always use it only in a well ventilated room 2 Do not attempt to use organic solvents to clean the microscope components other than the glass components To clean them use a lint free soft cloth s...

Страница 7: ...To remove an objective turn it counterclockwise while holding it with a rubber sheet etc so that your fingers don t slip Eyepieces Fixed at 10X Stage Condenser Microscope frame 4X 10X 40X 100X CX22LE...

Страница 8: ...ch the AC adapter and power cord as described in chapter 9 ASSEMBLY on page 17 For detailed description of each item refer to the page indicated inside parentheses NOTE FOR UNPACKING Power ON Power OF...

Страница 9: ...agm ring P 10 aFilters Diameter 45 mm P 5 Engage the required filters 5Revolving nosepiece P 10 6Coarse fine adjustment knobs P 8 Adjust the brightness 2Light intensity adjustment knob P 7 Items Used...

Страница 10: ...ront toward the rear 3 After sliding the specimen glass plates all the way return the bow shaped lever 3 gently 4 Rotating the upper knob which is the Y axis feed knob 4 moves the specimen in the vert...

Страница 11: ...obtained rotate the fine adjustment knob 2 to adjust to precise focus Working Distance WD The WD refers to the distance between each objective and the specimen when precise focus of the specimen is o...

Страница 12: ...d CAUTION 4 Adjusting the Interpupillary Distance Fig 11 The interpupillary distance adjustment is to regulate the two eyepieces according to that between your eyes so that you can observe a single mi...

Страница 13: ...epiece 1 so that the objective to be used comes exactly above the specimen Fig 12 Fig 13 5 Adjusting the Diopter Fig 12 The diopter adjustment is to compensate for the difference in eyesight between y...

Страница 14: ...bbles in the oil will affect the image quality make sure that the oil is free of bubbles To remove bubbles rotate the revolving nosepiece slightly to move the oil immersed objective by one reciprocati...

Страница 15: ...h both configurations above provide the same total magnification the higher magnification of the objective yields a better specimen resolution Field Number FN The field number is the diameter of the i...

Страница 16: ...ed in the immersion oil Remove the bubbles The specified immersion oil is not used Use the immersion oil supplied by Olympus 11 11 11 5 Part of image is defocused or image looks like it s flowing The...

Страница 17: ...ched to a higher magnification objective The specimen is upside down Set the specimen correctly with the cover glass on the top The cover glass is too thick Use a cover glass with thickness of 0 17 mm...

Страница 18: ...ustment range 48 to 75 mm 6 Stage Size 120 x 132 mm with mechanical stage Movement range 76 X axis x 30 Y axis mm Specimen holder Holds a single specimen 7 Condenser Type Abbe condenser N A 1 25 when...

Страница 19: ...age which is expressed in terms of the minimum distance between two points on the specimen surface Focal depth The depth range of a specimen in which focusing is obtained at a time Stopping down the a...

Страница 20: ...er to the input connec tor 3 on the rear of the microscope Do not tilt the microscope to the rear Doing so could damage the output connector 2 2 Connect the power cord s connector 4 to connector 5 fir...

Страница 21: ...ttom of the condenser until it clicks into place The darkfield ring enables darkfield observation using an objective from 4X to 40X OPTIONAL ACCESSORIES 10 1 System Diagram of Optional Accessories 10...

Страница 22: ...80 1 Fit the reflection mirror in the window lens of the microscope frame by aligning the mounting tab 2 The reflection mirror 1 should be pointed toward the bright area While observing the image thro...

Страница 23: ...observation 3 Screw the reticle holder containing the micrometer disk into the bottom of an eyepiece Be sure to screw in all the way by hooking your nail into the notch 3 on the holder at the end 4 Sc...

Страница 24: ...the cable 5 of the AC adapter into the hole of center part b of cord hanger Then bring it down so that it goes around the bottom part c of cord hanger and bring it into the hole b again Fig 26 4 Inser...

Страница 25: ...oscope when storing the microscope after use CAUTION Fig 28 Fig 29 5 Place the power cord on the cord hanger 4 as illustlated Fig 28 Removal 1 Insert the tool 9 such as screwdriver into the space betw...

Страница 26: ...1 Certified Cord A power cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The fittings are to b...

Страница 27: ...stalter sterreichischer Verband f r Elektrotechnik VE Danmarks Elektriske Materialkontroll DEMKO National Standards Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Elect...

Страница 28: ...18 20097 Hamburg Germany 3500 Corporate Parkway Center Valley Pennsylvania 18034 0610 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilby Road Mount Waverley VIC...

Отзывы: