17
Fig. 18
1
· Be sure to use the provided AC adapter. Use of any other AC
adapter may cause a malfunction.
· The power cord and AC adapter cord are vulnerable when
bent or twisted. Never subject them to excessive force.
· Make sure that the main switch
1
is set to “ ” (OFF) before
connecting the power cord.
· Always use the power cord provided by Olympus. If no power
cord is provided, please select the proper power cord by
referring to section “PROPER SELECTION OF THE POWER
CORD” at the end of this instruction manual.
1. Connect the output connector
2
of the AC adapter to the input connec-
tor
3
on the rear of the microscope.
Do not tilt the microscope to the rear. Doing so could damage
the output connector
2
.
2. Connect the power cord’s connector
4
to connector
5
firmly.
Be sure to supply power from a grounded, 3-conductor power
outlet using the proper power cord. If the power outlet is not
grounded properly, Olympus can no longer warrant the electrical
safety performance of the equipment.
3. Connect the power cord’s plug
6
to a wall power outlet
7
. (Fig. 18)
²
@
³
Fig. 17
|
5
Connecting the AC Adapter and
Power Cord
(Figs. 17 & 18)
ASSEMBLY
}Each standard set can be assembled by simply attaching the AC adapter and power cord.
CAUTION
CAUTION
CAUTION
*
Shapes of the power cord’s plugs and
power outlets are different by country.
7
6