CX21 MAINTENANCE MANUAL
REPAIR PROCEDURE
- 14 -
2. Mechanical Adjustment
2-1 Preparation for the tension adjustment of X/Y-wire
If a specimen image is moved when the stage is brought into the desired position of
specimen, it is necessary to adjust the wire tension of stage.
1)
2)
3)
OLYMPUS
CX21
7 5
7 0
6 0
5 0
1 0 0
1
1 0
1 2 0
1 3 0
140
150
1
7
160
0
10
20
30
(a)
Take off the mechanical stage (A) from the plane stage (B)
by removing the two screws (*1).
(A)
(B)
*1
arrow view (a): bottom
*1
(B)
(A)
*1: Fixing screws of mechanical stage
Screw: AB3X10SA (*1), 2 pcs.
Remove the specimen holder (C) from the
inner guide (D).
Screw: 3PSK2X4SA (*2), 3pcs.
(D)
*2
(C)
X-wire
* For the tension adjustment of Y-wire,
* For the tension adjustment of X-wire,
refer to next page.
refer to next page.
Remove the Y-cover (E).
Screw: CUTB3X8SA (*3), 3pcs.
*3
<indicated by arrows>
(E)
Y-wire