Olympus CelonLab ENT Скачать руководство пользователя страница 7

 

Safety Check Instruction

 

 

 

 

 

 

 

©  Celon AG 

Page 7 / 9 

 

Issue date: 01.09.2008 

SCI.991007-1.4 

Safety Check 

Instruction.doc 

No. 

Procedure 

 

Upper shut off limit 

240 

The following signals have to occur with a load resistor of 325 

Ω (normal function of the unit): 

-

 

The blue LED 

(signal lamp “power output”) glows. 

-

 

The red LED (signal 

lamp “resistance too low”) is dark. 

-

 

A high frequency continuous signal is audible. 

250 

The following signals have to occur with a load resistor of 450 

Ω (unit does 

not

 shut off): 

-

 

The blue LED 

(signal lamp “power output”) glows. 

-

 

The red LED (signal 

lamp “resistance too low”) is dark. 

-

 

A pulsed signal with a high audio frequency is audible (end of treatment). 

 

Documentation

 

260 

Document 

the test results in “SCR.991007”. 

 
 

5. High frequency leakage current (according to IEC 60601-2-2) 

No. 

Procedure 

 

Test equipment: 

Electrosurgical analyzer (Example: QA-ES, Metron) 

 

Load resistor 100 

Ω, 25 W (or more), 5 % tolerance, low inductive part, short time 

 

load 

 

Connection cables, HF-output (banana) to the electrosurgical analyzer and the load 

 

resistor 

 

Test accessories:  Power cord 

 

Footswitch 

270 

Measurement of the high frequency leakage current according to 19.3.101 a) 3) (bipolar application) as 
described in IEC 60601-2-2. The output being loaded and unloaded at rated load. 

Connect the unit via the power cord with a rated supply voltage. Connect the footswitch with the unit. 

280 

Connect the load resistor (100 

Ω) with the HF-output 1 and 2 and with the electrosurgical analyzer 

according to the analyzer’s instructions for use to measure high frequency leakage currents. Set the 
power level of the unit to “25” and measure the following parameter: 

Attention: Start measuring, if not already done, after setting the power level 

to “25”and make 

sure to release the footswitch before starting with a new measurement.

 

290 

High frequency leakage current of HF-output 1 at output load 100 

Ω: 

≤ 35.4 mA. 

300 

High frequency leakage current of HF-output 2 at output load 100 

Ω: 

≤ 35.4 mA. 

310 

Disconnect the load resistor (100 

Ω) from the HF-output 1 and 2. 

320 

Connect the HF-

output 1 and 2 with the electrosurgical analyzer according to the analyzer’s instructions 

for use to measu

re high frequency leakage currents. Set the power level of the unit to “25” and measure 

the following parameter: 

Attention: 

Start measuring, if not already done, after setting the power level to “25”and make 

sure to release the footswitch before starting with a new measurement.

 

330 

High frequency leakage current of HF-output 1 with open output: 

≤ 35.4 mA. 

340 

High frequency leakage current of HF-output 2 with open output: 

≤ 35.4 mA. 

 

Documentation

 

350 

Document 

the test results in “SCR.991007”. 

Содержание CelonLab ENT

Страница 1: ...maintenance take care of the different hardware versions which may be applicable see below The hardware version of the unit can be identified by the serial number which can be found on the type plate...

Страница 2: ...ountry medical device information Medizinproduktebuch 020 Labels present and legible Sample of symbol Unit is defibrillator safe Visible on the front panel see fig 2 Sample of symbol Caution Read the...

Страница 3: ...nnection for equipotential bonding Visible on the back of the unit see fig 1 Samples of label for power fuses Visible on the back of the unit see fig 1 Sample of symbol Non ionizing radiation Visible...

Страница 4: ...Celon AG Page 4 9 Issue date 01 09 2008 SCI 991007 1 4 Safety Check Instruction doc Fig 1 Sample of the backside of the CelonLab ENT power control unit Fig 2 Sample of the front panel of the CelonLab...

Страница 5: ...Documentation 110 Document the test results in SCR 991007 3 Power output No Procedure Test equipment Electrosurgical analyzer Example QA ES Metron Connection cables HF output banana to the electrosurg...

Страница 6: ...l analyzer according to the analyzer s instructions for use Set the power level of the unit to 12 and measure the following parameter Attention Start check procedure if not already done after setting...

Страница 7: ...loaded and unloaded at rated load Connect the unit via the power cord with a rated supply voltage Connect the footswitch with the unit 280 Connect the load resistor 100 with the HF output 1 and 2 and...

Страница 8: ...ety tester Example Unimet 1000 ST Bender Load resistor 100 25 W or more 5 tolerance low inductive part short time load Connection cables HF output banana to the load resistor Test accessories Power co...

Страница 9: ...c 8 Inspection label No Procedure 430 Cover or exchange an inspection label as shown below at the enclosure back of the unit and mark the due date of the next periodic safety check month year The unit...

Отзывы: