Olympus CelonLab ENT Скачать руководство пользователя страница 6

 

Safety Check Instruction

 

 

 

 

 

 

 

©  Celon AG 

Page 6 / 9 

 

Issue date: 01.09.2008 

SCI.991007-1.4 

Safety Check 

Instruction.doc 

 

 

Fig. 3: 

Sample of the front panel of the power control unit showing the HF-outputs. 

 
 
 
 

4. Valid resistance range 

No. 

Procedure 

 

Test equipment: 

Electrosurgical analyzer (Example: QA-ES, Metron) 

 

Connection cables, HF-output (banana) to the electrosurgical analyzer

 

 

Test accessories:  Power cord 

 

Footswitch 

200 

Connect the unit via the power cord with a rated supply voltage. Connect the footswitch with the unit. 

210 

Connect the unit via the connection cables at the HF-output 1 and 2 (see fig. 3) with the electrosurgical 
analyzer according to the analyzer’s instructions for use. Set the power level of the unit to “12” and 
measure the following parameter: 

Attention: Start check procedure, if not already done, after setting the power level 

to “12”and 

make sure to release the footswitch before starting with a new check procedure. 

 

Short circuit detection 

220 

The following signals and warnings have to occur at an output load of 10 

Ω (unit shut off): 

-

 

The 

blue LED (signal lamp “power output”) is dark. 

-

 

The red LED (signal 

lamp “resistance too low”) is blinking. 

-

 

An alarm signal (pulsed tone) is audible. 

230 

The following signals have to occur with a load resistor of 25 

Ω (normal function of the unit): 

-

 

The blue LED 

(signal lamp “power output”) glows. 

-

 

The red LED (signal lamp “resistance too low”) is dark. 

-

 

A continuous signal is audible. 

 
 
 
 
 
 
 

HF-output 1 

HF-output 2 

Содержание CelonLab ENT

Страница 1: ...maintenance take care of the different hardware versions which may be applicable see below The hardware version of the unit can be identified by the serial number which can be found on the type plate...

Страница 2: ...ountry medical device information Medizinproduktebuch 020 Labels present and legible Sample of symbol Unit is defibrillator safe Visible on the front panel see fig 2 Sample of symbol Caution Read the...

Страница 3: ...nnection for equipotential bonding Visible on the back of the unit see fig 1 Samples of label for power fuses Visible on the back of the unit see fig 1 Sample of symbol Non ionizing radiation Visible...

Страница 4: ...Celon AG Page 4 9 Issue date 01 09 2008 SCI 991007 1 4 Safety Check Instruction doc Fig 1 Sample of the backside of the CelonLab ENT power control unit Fig 2 Sample of the front panel of the CelonLab...

Страница 5: ...Documentation 110 Document the test results in SCR 991007 3 Power output No Procedure Test equipment Electrosurgical analyzer Example QA ES Metron Connection cables HF output banana to the electrosurg...

Страница 6: ...l analyzer according to the analyzer s instructions for use Set the power level of the unit to 12 and measure the following parameter Attention Start check procedure if not already done after setting...

Страница 7: ...loaded and unloaded at rated load Connect the unit via the power cord with a rated supply voltage Connect the footswitch with the unit 280 Connect the load resistor 100 with the HF output 1 and 2 and...

Страница 8: ...ety tester Example Unimet 1000 ST Bender Load resistor 100 25 W or more 5 tolerance low inductive part short time load Connection cables HF output banana to the load resistor Test accessories Power co...

Страница 9: ...c 8 Inspection label No Procedure 430 Cover or exchange an inspection label as shown below at the enclosure back of the unit and mark the due date of the next periodic safety check month year The unit...

Отзывы: