Problem
Cause
Remedy
Page
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
Interpupillary distance is incorrect.
Adjust interpupillary distance.
15
Incorrect diopter adjustment.
Adjust diopter.
15
Different eyepieces are used on left and
right.
Change on eyepiece to match the
other so that both sides are the same
type.
–
You are not accustomed to parallel opti-
cal axis.
When looking into eyepieces, do not
stare at image from the beginning but
see the overall field of view. It is some-
times recommended to turn your eyes
away from eyepieces, look far off and
look into eyepieces again.
–
5. Stage
a) Image shifts when you touch stage.
Stage is not properly mounted.
Clamp stage.
14
b) Stage cannot travel in Y-axis direc-
tion (U-SIC4R2/SIC4L2).
Y-axis travel is locked.
Release lock.
c) Stage knob is too light or too heavy
to rotate (U-SVRM/U-SVLM)
X-axis and/or Y-axis rotation tension is
not adjusted properly.
Adjust it again.
13
13