TROUBLESHOOTING GUIDE
Under certain conditions, performance of the scanning stage with ultrasonic may be adversely affected by factors other than
defects. If problems occur, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance.
Problem
Cause
Remedy
Page
1. Stage
a) The stage cannot be moved by
rotating the X-/Y-axis knobs.
Connection cable is disconnected.
Connect the cable.
12
The stage is moved until the stopper at
the end of stroke.
Use the area near the center of the
stage.
6
The error mode is entered because the
stage was moved quickly by hand.
Turn power off then on again.
6
b) The coarse/fine adjustment cannot
be switched by pressing the coarse/
fine selector button.
Connection cable is disconnected.
Connect the cable.
12
c) The stage cannot be moved smoothly
when a high-power objective is
used.
The operating environmental tempera-
ture has changed drastically.
Perform the auto compensation in
“System Setting” of the touch-panel
controller.
--
d) The stage movement has become
unsmooth after a long period of
use.
Fine particles produced due to wear
of the stage’s sliding surfaces are
deposited.
Set the control box to “ ” (OFF) and
move the stage manually into all of the
front, rear, left and right directions (full
strokes). And then, perform the Auto
Compensation of the stage
--
e) The auto compensation of the stage
fails.
The cable of the stage is stretched.
Loosen the cable and perform the auto
compensation again.
--
The auto compensation is not comple-
ted within the specified time.
Turn ON the power of each unit again
and perform the auto compensation
again. If you may fail in the auto com-
pensation for several times, contact the
Olympus sales office where you have
purchased the device.
--
f ) The stage movement range became
narrow.
The stage origin position is misreco-
gnized.
Compensate the stage.
7
2. Stage control box
a) The pilot lamp does not light.
The main switch of the control box is
not ON.
Set the main switch to “
I
” (ON).
6
Power cord is connected improperly.
Connect the power cord properly.
12
Fuse in control box is blown.
Replace the fuse with a new one.
6
Cables are connected improperly.
Connect the cables properly.
12