10
BX3-SSU
6
6
SPECIFICATIONS
Item
Specification
Scanning stage with ultrasonic
Applicable microscope frame BX63F, IX83P1ZF, IX83P2ZF, IX73P1F, IX73P2F, BX51WIF
Slide holder
Slide glass specimen x 2 (total thickness 0.9 to 1.2 mm, cover glass thickness
0.17 mm)
Stage center plate
Center plate with 25 mm hole (BX3-SSU accessory)
Center plate with 50 mm hole IX-CP50 (option)
Stage clips
IX2-SCL.
Stage drive stroke
X: 76 mm. Y: 52 mm.
Stage drive resolution: 0.2 µm
Maximum stage load
1 kg
Dimensions & weight
268(W) x 251(H) x 29(D) mm, approx. 3.2 kg.
XY controller
Stage control method
Rotary knobs (X-/Y-axis coaxial knobs), with coarse/fine selector button.
Displacement per turn
Coarse adjustment: 10 mm.
Fine adjustment: 0.2 mm.
Dimensions & weight
223(W) x 88(H) x 54(D) mm, approx. 1 kg
Control box
Rating
100-120/220-240 V 50/60 Hz 0.35/0.2 A
Front panel
Main switch (“
I
”: ON. “ ”: OFF), with pilot lamp LED.
Dimensions & weight
100(W) x 215(H) x 280(D) mm, approx. 3.9 kg
Operating environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104°F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly through 70% at 34°C
(93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation/Overvoltage category: II (in accordance with IEC60664-1)