background image

Complete Teardown, Cleaning, and Reassembly of the Olympus BHSU/BHTU Reverse Nosepiece Assembly

 

 

Revision 3

       

Page 11 of 19 

Place the turret assembly into a suitable work vise and 
use a heat gun to thoroughly heat the turret assembly, 
to soften the internal grease.  Then use a center punch 
or nailset tool to loosen the threaded retaining ring by 
placing  the  tip  of  the  tool  into  one  of  the  four  tooling 
holes in the threaded retaining ring (se

Figure 22

and 

driving the threaded retaining ring counter-clockwise by 
carefully tapping the tool with a small hammer or mallet 
(see 

Figure 23

).  Be careful and do not accidentally put 

the  center  punch  or  nailset  tool  into  one  of  the  five 
detent  notches  in  the  revolving  turret  (instead  of  into 
one of  the  four  tooling  holes  in  the  threaded  retaining 
ring),  or  the  revolving  turret  will  be  irreparably 
damaged when the punch is struck by the mallet. 

 

Figure 23 

 Loosen the threaded retaining ring

 

Continue  loosening  the  threaded  retaining  ring  and 
remove it, being careful that the perimeter bearing balls 
beneath the threaded retaining ring do not fall out and 
become lost in the process (see 

Figure 24

). 

 

Figure 24 

 Remove the threaded retaining ring

 

Remove the Perimeter Bearing Balls 

Carefully 

remove  the  3/32”  bearing  balls  (there  are  8

of  them)  from  the  channel  formed  by  the  outer 
perimeter  of  the  stationary  base  and  the  inner 

perimeter  of  the  revolving  turret.    A  small  magnet  can 
come in handy here (see 

Figure 25

).

 

 

Figure 25 

 Remove the perimeter bearing balls 

Remove the Stationary Base from the Turret 

Once  the  perimeter  bearing  balls  have  been  removed 
from  the  channel,  lift  the  stationary  base  free  of  the 
recess in the revolving turret and remove it (see 

Figure 

26

). 

 

Figure 26 

 Lift and remove the stationary base 

Thoroughly Clean the Various Components 

Use  a  suitable  solvent  (e.g.,  acetone)  to  thoroughly 
clean  all  the  old  grease  from  the  stationary  base, 
revolving  turret,  threaded  retaining  ring,  perimeter 
bearing balls, center-pivot ball, mechanical detent stop, 
pivot-adjustment  screw,  and  the  slotted  lock  ring,  in 
preparation for reassembly. 

Reinstall Stationary Base into the Revolving Turret 

Hold  the  stationary  base  such  that  the  center  pivot  is 
facing  downwards,  and  then  lower  the  stationary  base 
into the recess of the revolving turret (see 

Figure 27

).   

Содержание BHSU

Страница 1: ...this PDF document in its entirety for personal non commercial purposes only provided that the contents are not modified in any way including the copyright notices contained herein This document may be...

Страница 2: ...chanical Detent Stop 10 Remove the Black Knurled Rubber Grip Ring 10 Remove the Threaded Retaining Ring 10 Remove the Perimeter Bearing Balls 11 Remove the Stationary Base from the Turret 11 Thoroughl...

Страница 3: ...n the slotted lock ring 9 Figure 16 Remove the slotted lock ring 9 Figure 17 Remove the pivot adjustment screw 9 Figure 18 Remove the center pivot bearing ball 10 Figure 19 Remove the screws securing...

Страница 4: ...silicone RTV to revolving turret 15 Figure 45 Reinstall the knurled rubber grip ring 16 Figure 46 Remove any silicone RTV squeeze out 16 Figure 47 Place the wedge mount in position 16 Figure 48 Reinst...

Страница 5: ...he Olympus BHSU and BHTU microscope stands Note that the original Olympus service literature did not address the teardown and repair of the various revolving nosepiece assemblies used on BH 2 stands a...

Страница 6: ...st of solvents generally considered safe for nylon includes acetone diethyl ether heptane mineral spirits naphthalene and turpentine Before Starting with the Overhaul Before beginning the removal and...

Страница 7: ...ing the wedge mount to the stationary base of the turret assembly see Figure 6 It is important to remove these four screws before attempting to remove the corrective optics assembly to prevent the bar...

Страница 8: ...pivot adjustment screw unless you have access to such a tool since the slot in the lock ring will likely be damaged if you use an improper tool making removal of the lock ring much more difficult1 A...

Страница 9: ...g the knurled rubber grip ring on the outer perimeter of the revolving turret and use the slotted screwdriver with the notched tip see Figure 13 to loosen the slotted lock ring securing the pivot adju...

Страница 10: ...ent damaging the screw heads It might also be helpful to heat the screws with a heat gun before loosening them but do not melt the black knurled rubber grip ring in the process Figure 19 Remove the sc...

Страница 11: ...th the threaded retaining ring do not fall out and become lost in the process see Figure 24 Figure 24 Remove the threaded retaining ring Remove the Perimeter Bearing Balls Carefully remove the 3 32 be...

Страница 12: ...rease for the bearing balls Reinstall the Perimeter Bearing Balls Use tweezers to carefully set the 3 32 bearing balls there are 80 of them into the grease ring see Figure 29 placing the bearing balls...

Страница 13: ...at the center pivot bore has been greased reinstall the bearing ball into the freshly greased center pivot bore The grease will hold the center pivot bearing ball in the proper position during subsequ...

Страница 14: ...t with the other hand The motion of the turret should not feel gritty erratic or excessively stiff If it does loosen the slotted lock ring readjust the pivot adjustment screw and retighten the slotted...

Страница 15: ...o the stationary base see Figure 43 Figure 43 Secure protective cover using three screws Reinstall the Knurled Rubber Grip Ring The final assembly step for the turret assembly is to reinstall the blac...

Страница 16: ...sition While holding the wedge mount in this position place the threaded end of the corrective optics assembly through the center bore of the wedge mount and screw the corrective optics assembly into...

Страница 17: ...anical detents thereby maximizing the useful service life of the reverse nosepiece assembly Follow the procedures detailed in this document to remove the reverse nosepiece assembly from the microscope...

Страница 18: ...may be encountered if the threaded retaining ring has not been properly snugged down or if the center pivot adjustment screw has not been properly adjusted to remove the play in the center pivot mech...

Страница 19: ...on 5 40 4 Bearing balls stainless steel 1 4 G25 25 count various Amazon 4 95 5 Pliers soft jaw Non Scratch Pliers Micro Mark 85161 34 95 Pliers soft jaw Tamiya 74061 Amazon 32 64 6 Lens spanner tool p...

Отзывы: