24
Before use
Checking for mechanical defects
1. Check that the product has:
– no dents, cracks, kinks, or deformations
– no deep scratches
– no corrosion
– no missing or loose parts
– no cuts and other defects on the insulation
2. Check all markings on the product for clear visibility.
Checking for performed reprocessing
1. Make sure that the product has been properly reprocessed.
2. Visually inspect the product thoroughly. The product must be
visually clean.
3. Only use the product if it is visually clean. Reprocess the
product again if necessary.
Replace if necessary
If the instrument is damaged or does not function properly,
replace it. Contact an authorized service center for repair.
4.2 Attaching the neutral electrode
This section applies only to the following products:
A00010A, A00011A, A00012A, A0335.1, A0355, A0357, A0358,
A0391, A0392, A0393, O0120.1
●
Refer to the instructions for use of the electrosurgical
generator and the neutral electrode.
●
Attach the neutral electrode to the patient.
●
Connect the neutral electrode to the electrosurgical
generator.