manualshive.com logo in svg
background image

45MG — Getting Started Guide

5

2.

Couple the transducer on the thick sample of the test block.

3.

Press 

[CAL VEL]

.

4.

Once the thickness reading is stable, press 

[ENTER]

.

5.

Use the arrow keys to enter the known thickness.

6.

Press 

[CAL ZERO]

.

7.

Place couplant on the surface of the thin sample of the test block.

8.

Couple the transducer on the thin sample of the test block.

9.

Once the thickness reading is stable, press 

[ENTER].

10. Use the arrow keys to enter the known thickness.
11. Press 

[MEAS]

.

Note

: For more information about calibration, and about 

selecting the correct transducer for a specific application, refer to the 

45MG Ultrasonic Thickness 

Gage — User’s Manual

 (P/N: DMTA-10022-01EN), or contact Olympus.

Keypad Functions

The 45MG comes with either an English or international keypad. The functions on both keypads are 
identical. Many keys on the international keypad contain pictograms instead of text labels. In the 
45MG user documentation, keypad keys are referred to using the English label, displayed in bold and 
within brackets (ex.: 

[FILE]

).

Each key is labelled according to its primary function. The area immediately above certain keys 
indicates a secondary key function. The 

[

]

[

]

[

]

[

]

, and 

[ENTER]

 keys are used to select menu 

items and screen parameters, and to change parameter values. Use the 

[MEAS] 

key at any time to 

return to the measurement screen.

Transducer on a 
thick sample

Transducer on a

thin sample

English keypad

International keypad

Содержание 45MG

Страница 1: ...p the documentation CD in a safe accessible location Safety Signal Words Indicates an imminently hazardous situation calling attention to a procedure practice or the like which if not correctly performed or adhered to could result in death or serious personal injury Indicates a potentially hazardous situation calling attention to a procedure practice or the like which if not correctly performed or...

Страница 2: ... memory card USB communication connector Transmit Receive transducer connector 2 Transmit Receive transducer connector 1 To avoid the risk of electric shock do not touch the inner conductors of the T R 1 and T R 2 connectors Up to 200 V may be present on the inner conductor ...

Страница 3: ...al protective rubber boot if installed 4 Turn the battery door lock half a turn counterclockwise to the unlock position 5 Remove the battery compartment cover 6 Remove the batteries 7 Insert three new batteries into the battery compartment making sure to observe the correct polarity for each battery 8 Ensure that the gasket inside the battery compartment cover is clean and in good condition 9 Rein...

Страница 4: ...actory default settings for the transducers purchased The default settings use an approximate sound velocity for the stainless steel test block provided with the instrument 1 Connect the transducer to the transducer cable and then plug the cable into the T R 1 connector on the top of the instrument 2 Press to turn on the instrument 3 Press 2nd F FREEZE XDCR RECALL 4 In the menu select DEFAULT SING...

Страница 5: ...G Ultrasonic Thickness Gage User s Manual P N DMTA 10022 01EN or contact Olympus Keypad Functions The 45MG comes with either an English or international keypad The functions on both keypads are identical Many keys on the international keypad contain pictograms instead of text labels In the 45MG user documentation keypad keys are referred to using the English label displayed in bold and within brac...

Страница 6: ... battery USB or AC MicroSD memory card indicator Mode and update rate ID bar with Datalogger option Zoom indicator Rectification indicator Echo detection marker Delay value Mode MIN MAX and value Freeze indicator Alarm indicator Gain level Range value WAVE ADJ parameter Power indicator battery USB or AC MicroSD memory card indicator Mode and update rate ID bar with Datalogger option Previous thick...

Страница 7: ...de 7 Instrument Hardware Components Protective rubber boot optional Stand optional Strap rings at the four corners of optional rubber boot Color display Keypad I O door protecting USB connector Transducer connectors Power key ...

Страница 8: ...m Printed in the United Stated of America 2012 by Olympus All rights reserved All brands are trademarks or registered trademarks of their respective owners and third party entities DMTA 10024 01EN U8778520 Rev A September 2012 Printed on Rolland Hitech50 which contains 50 post consumer fiber 50 ...

Отзывы: