14 Measuring with the manual echo
‑
to
‑
echo detection mode
In severe corrosion situations where valid multiple echoes are not present, you must
use the standard mode to be able to measure thicknesses.
You can use dual element transducers in all three modes. You can use all
measurement, display, and datalogger functions with the echo-to-echo modes. The
internal datalogger recognizes and saves all necessary echo-to-echo information to
upload and download thickness, waveform, and setup data.
You do not have to switch between echo detection modes when measuring both
coated and uncoated areas because you can measure uncoated wall thicknesses using
an echo-to-echo mode.
To change the echo detection mode
1.
Press
[2nd F]
,
[ZOOM] (E
‑
TO
‑
E)
.
Dual element (DE)
manual echo-to-echo
(MEtoE) detection mode
E1 blank bar
Echo-to-echo detection
bar
[WAVE ADJ]
parameter
NOTE
TIP
Содержание 38DL PLUS
Страница 6: ...DMTA 10009 01EN U8778346 Rev D November 2016 Table of Contents vi...
Страница 8: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Abbreviations viii...
Страница 24: ...DMTA 10009 01EN U8778346 Rev D November 2016 Important Information Please Read Before Use 16...
Страница 26: ...DMTA 10009 01EN U8778346 Rev D November 2016 Introduction 18...
Страница 46: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 2 38...
Страница 56: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 3 48...
Страница 102: ...DMTA 10009 01EN U8778346 Rev D November 2016 Appendix 94...
Страница 106: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Tables 98...