DMTA-10009-01EN [U8778346], Rev. D, November 2016
Initial Setup
55
The available rectification modes are:
FULL
Shows the negative portion of the echo folded around the baseline so that both
positive and negative waveform lobes are displayed. Gives the best overall
representation of position and magnitude for most thickness measurement
applications.
FULL
is the default mode for dual element transducers.
HALF–
(
NEG
indicator)
Shows negative waveform lobes as positive and does not show the positive lobes.
HALF+
(
POS
indicator)
Shows positive waveform lobes and does not show the negative waveform lobes.
RF
Shows negative and positive lobes on either side of the baseline.
RF
is the default
mode for single element transducers.
4.4.4
Waveform Trace
The 38DL PLUS can display the waveform trace as a line (
OUTLINE
) or as a
FILLED
area (see Figure 4-6 on page 56). From the measurement screen, press
[DISPLAY]
to
access the
WAVEFORM TRACE
parameter.
A filled waveform trace is only possible when the waveform rectification is set to
FULL
,
HALF+
, or
HALF–
.
NOTE
Содержание 38DL PLUS
Страница 6: ...DMTA 10009 01EN U8778346 Rev D November 2016 Table of Contents vi...
Страница 8: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Abbreviations viii...
Страница 24: ...DMTA 10009 01EN U8778346 Rev D November 2016 Important Information Please Read Before Use 16...
Страница 26: ...DMTA 10009 01EN U8778346 Rev D November 2016 Introduction 18...
Страница 46: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 2 38...
Страница 56: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 3 48...
Страница 102: ...DMTA 10009 01EN U8778346 Rev D November 2016 Appendix 94...
Страница 106: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Tables 98...